background image

List-Magnetik GmbH 

EN FERRITE-CHECK 110 2022-11.doc 

16 

 

7.

 

T

ECHNICAL 

D

ATA

 

 

Application 

Measurement of ferrite content in austenitic and duplex steels 

Measuring range 

0,1 – 80 Fe% 
0,1 – 110 FN 

Measuring method 

Single measurement or continuous measurement with averaging 

Smallest measuring area 

ø 8 mm 

Resolution 

until 10: 0.01  
above 10: 0.1 

Ambient temperature 

0 – 50 °C 

Display 

illuminated high-contrast graphic OLED display 

Multilingual menu navigation 

German, English 

Power supply 

1x 1.5 V AA Mignon 

Operating time 

approx. 30 hours 

Dimensions 

Ø 28 x 94 mm 

Weight 

72 g (with battery) 

 

 

 

Summary of Contents for FERRITE-CHECK 110

Page 1: ...OPERATION MANUAL FERRITE CONTENT METER FERRITE CHECK 110 Firmware version 1 0 and up 2022 11...

Page 2: ...th the red button 3 4 Measurement display 4 5 Menu functions 5 a Off 5 b Calibrate 5 0 Base Calibration 6 FN Calibration 7 FN Value Adjust 10 c Mode 12 Average Measurement 12 Normal Measurement 12 d S...

Page 3: ...errite content Therefore the ferrite number is more commonly used Uneven or round surfaces as on rods as well as contamination of the surface can lead to incorrect or inaccurate measured values FERRIT...

Page 4: ...ignal tone confirms the measurement Switch off Press red key until display goes dark 3 OPERATING THE DEVICE WITH THE RED BUTTON Press the key briefly to scroll through the menu functions press and hol...

Page 5: ...measuring unit Setup Unit the measurement can be displayed in Fe When the measurement type is changed to the average value measurement Mode Average measurement the current single value and the average...

Page 6: ...f manually If you keep the key pressed for more than 3 seconds after switching off the firmware version is also displayed The automatic turnoff time is 2 minutes in the delivery state and can be chang...

Page 7: ...ut FN 140 FN 140 corresponds to 100 Fe and is the value of full iron or saturated ferrite material In the off hook state select in the menu 0 Base Calibration then place the instrument with the measur...

Page 8: ...on is normally performed on the calibration standard with the lower specified FN value Select FN Calibration in the off hook state then place the instrument with the measuring probe on the calibration...

Page 9: ...List Magnetik GmbH EN FERRITE CHECK 110 2022 11 doc 8 Function plan menu structure Arrow down red key short Arrow right red key long...

Page 10: ...List Magnetik GmbH EN FERRITE CHECK 110 2022 11 doc 9 Space for your notes...

Page 11: ...he case should you exchange it or use the right standard for two point calibration this value must be adjusted in the device The arrow to the right indicates that the FN values increase by pressing th...

Page 12: ...by briefly pressing the key again The exact FN value can then be set by pressing the key in both directions individually Press and hold the key to accept and save the FN value Because the internal ba...

Page 13: ...s are permanently recorded and arithmetically averaged In this way you can drag the device over a workpiece and see the current value as well as the average value of all previously measured values Aft...

Page 14: ...you can change language measuring unit and turnoff time LANGUAGE Selection of the menu language German or English UNIT Change the unit of measurement between FN and Fe The conversion from FN to Fe is...

Page 15: ...ay of the current battery voltage by long press TURNOFF TIME Selection of the automatic turnoff time of the device 2 minutes 15 minutes off the device is always switched on Off should only be selected...

Page 16: ...s the factory calibration can be reloaded here by performing a device reset The device switches off automatically 6 REPLACEMENT OF BATTERY As soon as the warning message Change batt appears when the d...

Page 17: ...N Measuring method Single measurement or continuous measurement with averaging Smallest measuring area 8 mm Resolution until 10 0 01 above 10 0 1 Ambient temperature 0 50 C Display illuminated high co...

Page 18: ...ic Field Meters Magnetic Permeability Meters Magnetizing and Demagnetizing Equipment We advise and provide tailor made solutions for your specialized requirements in magnetizing demagnetizing and meas...

Reviews: