iv | Safety Information
Power Cord
Use the supplied hospital grade power cord with the correct plug for your power source.
• The power cord is the only recognized disconnect device for this product. To power off the product,
disconnect the power cord from the AC mains.
• The product and other medical equipment should be positioned so that the power cord and connection to AC
mains is readily accessible.
• If an extension cord or power strip is needed for connection of this product to AC mains, confirm that the
power cord plug can be securely connected to the cord or power strip.
• This product should be powered from a center tapped circuit when used in the US at voltages over 120 VAC.
Grounding
It is the responsibility of the installer to ensure that the equipment is installed in accordance with applicable
hospital, local and national electrical codes.
An equipotentiality post, located on the back of the equipment, may be used for the purpose of bonding the
VIMA chassis to other equipment to ensure that all devices are at the same potential. Any such bond must be
installed in accordance with applicable electrical codes. The equipotentiality (ground) post is shown on page 4.
Data Storage
This product is designed to provide temporary data storage. Completed procedures and all corresponding data
should be exported, or other otherwise transferred, at intervals to prevent VIMA from reaching 75% capacity.
A notification is provided when the storage utilization reaches 75%. When storage utilization reaches 95%, an
alert is provided and recording and image capture features are blocked.
It is recommended to backup all stored data.
Recycling
Follow local governing ordinances and recycling plans regarding the recycling or disposal of this
product.