Troubleshooting and Test Section
8
Problem
Possible Causes and Remedies
Power switch does not
illuminate in the
ON
state
Loose power cable: Verify that the power cable is fully inserted into the unit’s
power connector and that the power supply’s connector collar is properly
threaded onto the
ScaleOR
’s power connector.
Defective power supply: Replace the power supply.
Wall socket: Some wall sockets have on / off switches built in. If the socket you
are using has a built in switch, verify that is in the on position.
Mitigation:
Bypass the
ScaleOR
by connecting the output of the video source directly to a
monitor.
Confidence test
Turn the
ScaleOR
off, then connect its
DVI
output port to a powered up
monitor. Turn the
ScaleOR
on, the
NDS
logo should appear on the monitor
approximately 10 seconds after the ScaleOR is turned on. Turn the
ScaleOR
off,
then connect its
SDI
output port to the monitor and repeat the test. If your
ScaleOR
has a
Fiber Optic
output, turn the
ScaleOR
off, then connect its
fiber
Optic
output port to the monitor and repeat the test.
Video is not displayed
If the
ScaleOR
’s power button’s blue ring is flashing once per second, then the
unit is not detecting a video signal. Verify that an input signal is connected to
the unit’s input. If there is a connection try replacing the video cable. If video
appears, then the cable is bad. If video does not appear, then connect the unit
to a different video source. If video appears, then the previous video source is
turned off, or defective. If video still does not appear, then the
ScaleOR
’s Input
module may be defective, or not properly seated. Try reseating the input
module. If that doesn’t clear the problem then the input module is defective.
If you determine that the input module is defective, please contact customer
service for assistance.
Mitigation:
If possible, connect the video source directly to the video display.
Video is not displayed
from a given output port
Verify that video is displayed from the
DVI
or
SDI
output port. If video from the
DVI
port is displayed, and not from the
SDI
port then
ScaleOR
’s output mode
may set to a
DVI
only mode (see
Output Modes Table
on page 15), the cable
connecting the
SDI
port to a monitor may be bad or the
SDI
output is defective.
If video from the
SDI
port is displayed, and not from the
DVI
port then the cable
from the
DVI
port may be bad, or the
DVI
port is defective.
If you determine that one of the output ports is defective, please contact
customer service for assistance.
External control failure
USB cable is loose or defective: Verify that the USB cable is fully inserted into the
ScaleOR
’s USB port and the PC’s USB port.
Control program has failed: Restart the program
Controlling PC has failed: Restart the PC
Mitigation:
Control the
ScaleOR
from its front panel Keypad
17