background image

NAVAIR 17-20AQ-347 

SECTION 1 

INTRODUCTION AND DESCRIPTION 

1.1  This procedure describes the calibration of the Wavetek 1271 Digital Multimeter with or without OPT12 or 20.  The 

instrument being calibrated is referred to herein as the TI (Test Instrument).    

1.2  All comments concerning this procedure should be directed to the Measurement Science Department, Corona Division, 

Naval Surface Warfare Center, P.O. Box 5000, Corona, CA 92878-5000. 

1.3  This procedure includes tests of essential performance parameters only.  Any malfunction noticed during calibration, 

whether specifically tested for or not, should be corrected. 

1.4  The TI is not calibrated to 1 MHz above 10 V and not calibrated over 500 V.  It is calibrated using the manufacturer’s 

12 month specifications.  This is an NSWC Measurement Science Department (MSD) assigned tolerance. 

1.5  It should be noted that the specifications cited below, in Table 1, are for the Fluke 8505/8506 series digital multimeters 

(unless noted otherwise).  Since the Fluke specifications were the “procurement specifications” for the Wavetek 1271 (TI), a 
SPECIAL CALIBRATION tag shall be affixed to the TI, upon completion of calibration, indicating that the TI was calibrated to 
the (12 month) Fluke 8505/8506 specifications.  This is an NSWC/MSD assigned specification. 

Table 1.  Calibration Description 

TI 

Characteristics 

Performance 

Specifications 

Test 

Method 

Direct voltage 

Ranges:  100 mV, 1, 10, 100 and 1000 V dc 
Tolerances: 

100 mV:  

±

(0.00403% iv + 90 counts) 

1 V:  

±

(0.0024% iv + 9 counts) 

10 V:  

±

(0.0019% iv + 9 counts) 

100, 100 V:  

±

(0.00297% iv + 9 counts) 

Tested with a voltage 

measurement system 
containing a voltage divider, 
reference divider, and DC 
voltage standard. 

Resistance 

Ranges:  10, 100 

Ω

, 1, 10, 100 k

Ω

, 1, 10, 

100 M

Ω

 

Tolerances: 

10 

Ω

:  

±

(100 ppm + 20 counts) 

100 

Ω

:  

±

(60 ppm + 1.4 counts) 

1, 10, 100 k

Ω

, 1 M

Ω

:   

   

±

(60 ppm + 0.8 counts) 

10 M

Ω

:  

±

(398 ppm + 0.8 counts) 

100 M

Ω

:  

±

(1004 ppm + 1 count) 

Compared with a known 

resistance. 

AC RMS Voltage 

Tolerances: 

100 mV range: 

10 to 40 Hz:  

±

0.152% iv 

40 to 20 kHz:  

±

(0.035% iv + 5 

μ

V) 

20 to 50 kHz:  

±

0.0825% iv 

50 to 100 kHz:  

±

0.308% iv 

Compared with known direct 

voltages by thermal voltage 
conversion. 

 

10 V range: 

10 to 40 Hz:  

±

0.152% iv 

40 to 20 kHz:  

±

0.025% iv 

20 to 50 kHz:  

±

0.0825% iv 

50 to 100 kHz:  

±

0.308% iv 

100 to 200 kHz:  

±

0.689% iv 

200 to 500 kHz:  

±

2.04% iv 

500 kHz to 1 MHz:  

±

4.49% iv 

 

Summary of Contents for 17-20AQ-347

Page 1: ...N OF COMMANDER NAVAL AIR SYSTEMS COMMAND DISTRIBUTION STATEMENT C DISTRIBUTION AUTHORIZED TO U S GOVERNMENT AGENCIES AND THEIR CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMI...

Page 2: ...he document NOTE On a revised page with the exception of the Title the A and the i pages the technical changes are indicated by a vertical line in the outer margin of the page Page No Revision Page No...

Page 3: ...0 4 3 Resistance Tests Option 20 Only 15 4 4 Calibration Factors 16 4 5 Securing The TI 17 Navy Calibration Checklist 19 ILLUSTRATIONS Figure Page 1 DC Volts Reference Setup 20 V or less 5 2 DC Voltag...

Page 4: ...curement specifications for the Wavetek 1271 TI a SPECIAL CALIBRATION tag shall be affixed to the TI upon completion of calibration indicating that the TI was calibrated to the 12 month Fluke 8505 850...

Page 5: ...100 kHz 0 308 iv 1000 V range 10 to 40 Hz iv 600 V x 02 0 152 0 2 in 40 Hz to 20 kHz iv 600 V x 02 0 025 0 2 in 20 to 50 kHz iv 600 V x 02 0 0825 0 2 in 50 to 100 kHz iv 600 V x 02 0 308 0 2 in NOTES...

Page 6: ...1000 V Uncertainties 100 mV 0 003 iv 1 V 0 000825 iv at 1 V 10 V 6 ppm at 19 V 100 V 0 001 iv at 100 V 1000 V 0 001 iv at 1000 V Fluke 7105A Required Components 2 1 1 DV source Fluke 332B or 335A 2 1...

Page 7: ...auxiliary equipment controls as necessary to avoid damage to the equipment and so that dangerous voltages will not be present on output terminals when the power switches are turned on 3 2 Ensure that...

Page 8: ...m up for 4 hours Ensure that the TI CAL switch TI rear panel is set to DISABLE DV SOURCE NULL DETECTOR VOLTAGE DIVIDER HI LO GND ERS 1 1 1 0 GUARD OUTPUT Figure 1 DC Volts Reference Setup 20 V or less...

Page 9: ...isted the CF value is 100 0 001 0 01 0 001000 V if the AC DC difference is a 0 001 the CF value would be 100 0 001 0 01 0 001000 V 3 8 Set all other calibration equipment controls as necessary to avoi...

Page 10: ...on is within the tolerance limits listed in the following table TI RANGE Nominal V Tolerance Limits V 100 mV 0 0000 3 0 0090 3 to 0 0090 3 1 V 0 000000 0 000009 to 0 000009 10 V 0 00000 0 00009 to 0 0...

Page 11: ...ithin the tolerance limits listed in the following table TI RANGE Divider Setting Nominal V Tolerance Limits V 100 mV 0 0005000 10 0000 3 9 9906 3 to 10 0094 3 100 mV 0 0050000 100 0000 3 99 9869 3 to...

Page 12: ...set the reference divider standard cell dials for the value of the 1 volt reference 4 1 8 Zero the null detector on its 10 V range then set its OPR ZERO switch to OPR Set the DV source power switch to...

Page 13: ...ntain a null on the null detector 10 V range taking precautions as required to avoid meter pegging TI RANGE V Input V Output V DV Source V Nominal V Tolerance Limits V 100 100 100 100 000 100 000 99 9...

Page 14: ...hat is applicable to the frequency of the measurement listed in the following table For example if the measurement frequency is 20 kHz and the CF at 20 kHz from step 3 6 is 0 000030 set the DV source...

Page 15: ...to open sensitivity switch to low and polarity switch to the opposite setting as in step 4 2 3 3 NOTE If the TTS polarity switch has been disconnected skip to step 4 2 3 12 4 2 3 8 Repeat steps 4 2 3...

Page 16: ...and the sensitivity switch to low 4 2 6 Set the TTS mode switch to AC search select the correct range so that the percent input meter indicates in the green area and then switch the mode switch to th...

Page 17: ...to the frequency of measurement as obtained in step 4 2 3 8 4 2 11 2 Verify that the TI indication is within the tolerance limits listed in the following table AV Source TI RANGE mV Output Volts Freq...

Page 18: ...ton Select the Resl function Select 5 digit resolution Depress the TI OHMS pushbutton 4 3 3 Connect a short across the 10 standard resistor 4 3 4 Depress the TI ZERO pushbutton Wait until the TI BUSY...

Page 19: ...out of tolerance 4 4 2 To begin setting the calibration factors engage the CAL switch on the rear panel of the TI 4 4 3 Perform an INTERNAL CHARACTERIZATION This is accomplished by performing the foll...

Page 20: ...ted at this point Use the TI SECOND RUN checklist column to repeat the necessary steps and verify that these steps are now within the given tolerance limits 4 5 2 Affix a SPECIAL CALIBRATION label to...

Page 21: ...0 009 4 1 4 Range Divider V V 100 mV 0 0005000 10 0000 3 9 9906 3 to 10 0094 3 0 0050000 100 0000 3 99 9869 3 to 100 0131 3 1 V 0 0050000 0 100000 0 099989 to 0 100011 0 0500000 1 000000 0 999967 to...

Page 22: ...to 1 00308 1 MHz 1 00000 0 95510 to 1 04490 4 2 4 10 V 10 Hz 10 0000 9 9848 to 10 0152 50 Hz 10 0000 9 9975 to 10 0025 1 kHz 10 0000 9 9975 to 10 0025 20 kHz 10 0000 9 9975 to 10 0025 50 kHz 10 0000 9...

Page 23: ...960 3 to 100 040 3 20 kHz 100 000 3 99 960 3 to 100 040 3 50 kHz 100 000 3 99 918 3 to 100 082 3 100 kHz 100 000 3 99 692 3 to 100 308 3 4 3 Resistance Tests Option 20 only 4 3 6 10 range ref 10 0000...

Reviews: