Xltek EMU40EX Battery Pack 

 

User Manual

 

 

 

 

USING THE EMU40EX HEADBOX WITH THE BATTERY 
PACK 

 

Connect Battery Pack (Fig.1) to EMU40EX Breakout Box using EMU40EX power cable (Fig.4). 

 

Green light

 

(“Battery”) will keep flashing as long as the battery is at least 20% full. 

 

When the Battery Pack is discharged to less than 20% of its capacity, flashing light “Battery” changes 
its color to 

RED

. This is a warning for you to complete the study or to change the power source in less 

than 1 hour. 

 

Even after the Battery Pack is discharged and all lights are off, there may still be some power available 
stored in the internal battery of the EMU40EX Breakout Box. 

USEFUL TIPS 

 

In order to check condition of the battery simply connect the EMU40EX Power Cable (Fig.4). There is 
no need to connect the headbox for this check. Three cases are possible: 
1. “Battery” LED is off 

the battery is completely discharged. 

2. “Battery” LED is flashing GREEN  the battery is more than 20% full. 
3. “Battery” LED is flashing RED 

the battery is less than 20% full. 

 

Please do not leave  EMU40EX Power Cable (Fig.4) plugged  if  you do  not  intend to use  the  Battery 
Pack for a long time. When the EMU40EX Power Cable is unplugged, the Battery Pack automatically 
enters low-power standby mode prolonging battery life. 

 

All lithium batteries are “perishable products”. Their life expectancy is limited to 500 charge-discharge 
cycles or 5 years due to chemical degradation of active electrodes. You may see very slow degradation 
of the lithium cell performance even if it is not used. 

 

Lithium battery will last much longer when using more 

shallow charge-discharge cycles

 rather than 

full-charge 

— full-discharge cycles.

 Battery life in charge-discharge cycles is specified for the latter 

(worst) case. Use fresh or fully charged batteries before beginning a study. 

 

If you are not going to use Battery Pack for a few weeks or longer, 

please pre-charge it approximately 

to 50~75%.

 This is optimal for long-term storage. Do not store a totally discharged battery as this may 

render the battery pack inoperable.  

 

Do  not  continuously  leave  the  battery  pack  charging.  Overcharged  cell  leads  to  faster  electrode 
degradation.  

 

If you need longer continuous operation two (or more) batteries can be used to prolong disconnected 
operation  of  EMU40EX  headbox.  This  is  limited  by  the  memory  storage  capacity  of  the  EMU40EX 
headbox. 

Summary of Contents for Xltek EMU40EX

Page 1: ...EMU40EX Battery Pack User Manual...

Page 2: ...or in part except with the prior written permission of Natus Neurology Incorporated The copyright and the foregoing restrictions on the copyright use extend to all media in which this information is p...

Page 3: ...D USERS AND PATIENT TARGET GROUP 7 CLINICAL BENEFITS 7 CONTRAINDICATIONS AND SIDE EFFECTS 7 CHARGING THE BATTERY PACK 7 USING THE EMU40EX HEADBOX WITH THE BATTERY PACK 8 USEFUL TIPS 8 MAINTENANCE 9 SP...

Page 4: ...ording to IEC 529 The Battery Pack is not water proof Prevent detergent solution or cold sterilization agents from seeping into the electronics of the system Battery pack stored by plugging into the c...

Page 5: ...to 50 75 before storage i e charge it for about 2 hours Device dropped or damaged in transit use could lead to loss of function or delayed diagnosis Inspect the device prior to each use and do not use...

Page 6: ...Xltek EMU40EX Battery Pack User Manual 6 Fig 1 Battery Pack Fig 2 USB Power Adapter Fig 3 USB power Cable Fig 4 EMU40EX Power Cable...

Page 7: ...The EMU40EX batter pack extends the period that a patient may be in an ambulatory position or not physically connected to an amplifier base through a cable to facilitate the continuous recording of a...

Page 8: ...u do not intend to use the Battery Pack for a long time When the EMU40EX Power Cable is unplugged the Battery Pack automatically enters low power standby mode prolonging battery life All lithium batte...

Page 9: ...acity 2500 2700mAh Energy capacity 9 8Watt hours Input charging voltage 5 12VDC Charging current limited to 500 1000mA 4 2V Charging time 3 hrs Charge retention storage 80 at Humidity 65 20RH 20 to 20...

Page 10: ...ons state that electrical and electronic waste must be separately collected for the proper treatment and recovery to ensure that WEEE is reused or recycled safely In line with that commitment Natus ma...

Page 11: ...State in which the user and or patient is established Refer to the Natus website for an electronic copy of this document INSTRUCTIONS TO ACCESS THE eIFU A copy of the Instructions for Use in PDF form...

Page 12: ...ation to be supplied Manufacturer Indicates the medical device manufacturer ISO 15223 1 Symbol 5 1 2 Medical devices Symbols to be used with medical device labels labelling and information to be suppl...

Page 13: ...1 Table D 2 2 Medical electrical equipment Part 1 General requirements for basic safety and Essential performance General warning sign Indicates a hazard of potential personal injury to patient or op...

Page 14: ...Website natus com 033980 Rev 1 03 2021 A Total Service Solution Natus Systems are backed by a comprehensive extendable warranty Our support team is available around the clock Our technical staff prov...

Reviews: