5.4
The TEOAE test
5.4.1
TEOAE test protocol
Protocol 8 measures non-linear TEOAE. The frequency range for the test signal is approximately 1 kHz to 4 kHz.
A
Pass/Clear Response
result requires detection of 8 peaks in the range 6 ms to 12 ms of the response.
5.4.2
Starting the test
1. To start the test, press the
Start
button that is next to
Protocol 8
, and that cor-
responds to the ear in which you placed the probe.
Calibration is performed before the actual test starts.
The test starts automatically when calibration is completed.
If the calibration fails due to large ear canal volume (for example, due to a PE tube in
the patient's ear), and if the
Manual start
option is enabled in
Test Settings
, then a
double arrow appears on the result screen. Press the double-arrow to start the meas-
urement with maximum calibrated output levels.
34
Otometrics - MADSEN Alpha OAE+
5 Testing with Alpha OAE+