
Revised 11/21/18
4-7
Electromagnetic Compatibility (EMC)
Table 8 – Immunity Test Levels - Signal Input / Output Parts Port
Phenomenon
Basic EMC Standard
Immunity Test Levels - Professional Healthcare
Facility Environment
Electrostatic Discharge
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Electrical Fast Transients/Bursts
IEC 61000-4-4
± 1 kV
100 kHz repetition frequency
Surges Line-to-Ground
(Common Mode)
IEC 61000-4-5
± 2 kV
Conducted Disturbances Induced
by RF Fields
IEC 61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM bands between 0,15 MHz and 80 MHz
80% AM at 1 kHz
Summary of Contents for Nicolet
Page 2: ......
Page 6: ...Nicolet Cortical Stimulator 4 Revised 11 21 18 Blank page ...
Page 11: ...Revised 11 21 18 1 1 1 General Information ...
Page 12: ...Nicolet Cortical Stimulator 1 2 Revised 11 21 18 Blank Page ...
Page 23: ...Revised 11 21 18 2 1 2 Nicolet Cortical Stimulator Overview ...
Page 24: ...Nicolet Cortical Stimulator 2 2 Revised 11 21 18 Blank Page ...
Page 41: ...Revised 11 21 18 2 19 Overview Blank page ...
Page 42: ...Nicolet Cortical Stimulator 2 20 Revised 11 21 18 ...
Page 43: ...Revised 11 21 18 3 1 3 Performing a Cortical Stimulation Study ...
Page 44: ...Nicolet Cortical Stimulator 3 2 Revised 11 21 18 Blank Page ...
Page 69: ...Revised 11 21 18 4 1 4 ElectromagneticCompatibility EMC ...
Page 70: ...Nicolet Cortical Stimulator 4 2 Revised 11 21 18 Blank Page ...
Page 77: ...Revised 11 21 18 5 1 5 Error Codes ...
Page 78: ...Nicolet Cortical Stimulator 5 2 Revised 11 21 18 Blank Page ...