Revised 11/21/18
3-17
Performing a Cortical Stimulation Study
6.
Pulse Freq
- To check the Pulse Frequency, click on the
Pulse Freq.
show
menu button and then select the desired pulse frequency rate (50 Hz,
typical). The last selected setting becomes the startup default setting.
7.
Pulse Duration
- To set the Pulse Duration, click on the
Pulse Dur.
show
menu button and then select the desired pulse duration:
100 - 1000 uSec
.
The last selected Duration setting becomes the startup default setting.
8.
Stimulation level
- At startup, the Stimulation level is set to 0.0 mA. Click
on the
Set Stim
show menu button and then select the desired stimulation
level:
0 - 15 mA
(8 mA or less is typical). If any parameter is changed (other
than the Stimulation level), the stimulation level will reset to 0.0 mA.
9. Begin with a lowest current level and increase in small increments.
10. If you want to mark the channels with ID tags prior to and after stimulation,
press the
Mark
Channel
button.
11.
Check Stim
- To check the Stimulator Integrity, click on the
Check
Stim
button to check the integrity of the stimulator without stimulating the
patient.
Summary of Contents for Nicolet
Page 2: ......
Page 6: ...Nicolet Cortical Stimulator 4 Revised 11 21 18 Blank page ...
Page 11: ...Revised 11 21 18 1 1 1 General Information ...
Page 12: ...Nicolet Cortical Stimulator 1 2 Revised 11 21 18 Blank Page ...
Page 23: ...Revised 11 21 18 2 1 2 Nicolet Cortical Stimulator Overview ...
Page 24: ...Nicolet Cortical Stimulator 2 2 Revised 11 21 18 Blank Page ...
Page 41: ...Revised 11 21 18 2 19 Overview Blank page ...
Page 42: ...Nicolet Cortical Stimulator 2 20 Revised 11 21 18 ...
Page 43: ...Revised 11 21 18 3 1 3 Performing a Cortical Stimulation Study ...
Page 44: ...Nicolet Cortical Stimulator 3 2 Revised 11 21 18 Blank Page ...
Page 69: ...Revised 11 21 18 4 1 4 ElectromagneticCompatibility EMC ...
Page 70: ...Nicolet Cortical Stimulator 4 2 Revised 11 21 18 Blank Page ...
Page 77: ...Revised 11 21 18 5 1 5 Error Codes ...
Page 78: ...Nicolet Cortical Stimulator 5 2 Revised 11 21 18 Blank Page ...