background image

 

 

neoBLUE

®

 mini System User Manual  

 P/N 051440D 

 

 

 

 

 

Warning!  Attachment of Arm with Pole Mount:

 When attaching the light 

to any floor standing pole, the user must validate that EN 60601-1/UL2601-1 

standards for stability in normal use is met. 

 

 

 

Warning!  Attachment of Arm with Pole Mount:

 Ensure that the pole 

mount is properly seated around the pole prior to use to prevent slippage during 

use.  Ensure that all screws and bolts are secure prior to use. 

 

 

 

Warning!  Weight Limitations of Arm:

 The arm has been validated to hold 

the weight of the light enclosure only.  Do not hang anything from the arm or 

power cord. 

 

 

 

Warning! Arm Adjustments:

 The arm folds into itself for optimal space 

utilization.  Use care when repositioning the arm to prevent pinching yourself. 

 

 

 

Warning!  Eye Protection:

 Do not look directly into the LEDs. During 

treatment, always protect the baby’s eyes with eye patches or equivalent. 

Periodically and/or per your hospital protocol, verify that the baby’s eyes are 

protected and free of infection. Patients adjacent to the light may also need to be 

protected with eye patches or equivalent.  

 

 

 

Warning!

  

Skin Temperature:

 The use of skin-controlled mode of the 

incubator or radiant warmer is recommended. In addition, use of reflective foils 

may cause hazardous body temperatures. Monitor the infant’s skin temperature per 

your hospital policy during phototherapy to avoid fluctuations in body 

temperature.    

 

 

 

Warning

!

  

Heat Supply: 

The phototherapy light may impact the heat supply 

in thermotherapy devices (incubators, radiant warmers, or heated mattresses) and 

the patient’s body temperature. 

 

 

 

Warning!  Use with Radiant Warmers:  

Do not place the phototherapy light 

directly under radiant heat sources. The light should be placed up to 12 inches 

(30.5 cm) from the infant, in line with the side of the heat element and angled 

toward the infant.

 

 

 

 

Warning!

  

Ambient Conditions:

 Varying ambient conditions, such as the 

ambient temperature and/or different radiation sources, may adversely affect the 

patient. Please refer to your hospital phototherapy policy and procedure regarding 

appropriate ambient conditions. 

 

 

 

Warning!

  

Operator Safety:

 Sensitive individuals may experience headache, 

nausea or mild vertigo if he/she stays too long in the illuminated area.  Using the 

neoBLUE mini device in a well-lighted area or wearing glasses with yellow lenses 

can alleviate potential effects.  Guard Dog Bones glasses (p/n 413BB) are 

recommended and are available online at 

www.safetyglasses.com

 or via phone at 1-

800-870-6189 (US Only). 

 

 

 

Warning!

  

Photoisomers:

 Bilirubin Photoisomers may cause toxic effects. 

Summary of Contents for neoBLUEmini

Page 1: ...____________________________________________________________ User Manual...

Page 2: ...ay not be reprinted or copied in whole or in part without written consent from Natus Medical Inc The content of this manual may change without notice Natus Medical Inc 1501 Industrial Road San Carlos...

Page 3: ...2 General Safety Information 2 3 Safety Symbols 3 Components and User Controls 3 1 Components 3 2 User Controls 4 Assembly and Operating Instructions 4 1 Setting Up the Device 4 2 Preparing Infant fo...

Page 4: ...bassinets or radiant warmers The light output is optimized to provide a peak intensity of 30 35 W cm2 nm Blue LEDs emit light in the range of 400 550 nm peak wavelength 450 470 nm This range correspon...

Page 5: ...present danger or possible injury to the patient and or user Important An instruction provided to help ensure correct clinical results and provide quality assurance to the phototherapy procedures Cau...

Page 6: ...er is recommended In addition use of reflective foils may cause hazardous body temperatures Monitor the infant s skin temperature per your hospital policy during phototherapy to avoid fluctuations in...

Page 7: ...ous oxide or other anesthetic agents Warning Disconnect Electrical Power Always switch off the power and disconnect the power cord when cleaning the light Important Use of Nonstandard Components The u...

Page 8: ...neoBLUE mini System User Manual P N 051440D 5 Natus Manufacturer WEEE Directive Waste from Electrical and Electronic Equipment Directive Symbol Product should not be disposed of in normal waste...

Page 9: ...al safety by protecting the baby and the unit from incidental debris or fluids Arm The adjustable arm allows positioning of the light in multiple locations and angles The arm is adjustable in five loc...

Page 10: ...se check the intensity of the light using a spectrophotometer per your institution s procedures see Section 6 1 Checking the Light Intensity The light has been calibrated with the Ohmeda BiliBlanket M...

Page 11: ...inches 45 7 cm 22 25 W cm2 nm 48 square inches 309 7 cm Approximately 6 x 8 15 2 cm x 20 3 cm 4 3 Administering Phototherapy 1 Switch on power using the power switch Warning Operator Safety Sensitive...

Page 12: ...122 F Fan not working Verify that vents are not blocked If the lens surface has become hotter than 50 C 122 F verify that the fan is functioning correctly Baby becomes too warm Fan not working Verify...

Page 13: ...The light has been calibrated with the Ohmeda BiliBlanket Meter II to deliver a peak intensity of 30 35 W cm2 nm at a 12 inch 30 5 cm distance from the baby Please refer to section 9 of this manual or...

Page 14: ...eter The intensity of the light is set at the factory to 30 35 W cm2 nm at a 12 inch 30 5 cm distance from the light enclosure This measurement is taken at the central area of the effective surface ar...

Page 15: ...12 in over 6 hrs 18 F 10 C warmer than ambient at 30 5 cm over 6 hrs Electrical Mains 85 264 VAC 47 to 63 Hz Maximum Rating 0 7A 100 240V 50 60 Hz Safety Leakage current 100 A Audible Noise 60 dB Dim...

Page 16: ...tem can be placed closer to the infant yielding a higher intensity or higher dosage treatment To adjust the neoBLUE mini light to the desired intensity when changing the distance please refer to the s...

Reviews: