User Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  

Summary of Contents for neoBLUEmini

Page 1: ...____________________________________________________________ User Manual...

Page 2: ...ay not be reprinted or copied in whole or in part without written consent from Natus Medical Inc The content of this manual may change without notice Natus Medical Inc 1501 Industrial Road San Carlos...

Page 3: ...2 General Safety Information 2 3 Safety Symbols 3 Components and User Controls 3 1 Components 3 2 User Controls 4 Assembly and Operating Instructions 4 1 Setting Up the Device 4 2 Preparing Infant fo...

Page 4: ...bassinets or radiant warmers The light output is optimized to provide a peak intensity of 30 35 W cm2 nm Blue LEDs emit light in the range of 400 550 nm peak wavelength 450 470 nm This range correspon...

Page 5: ...present danger or possible injury to the patient and or user Important An instruction provided to help ensure correct clinical results and provide quality assurance to the phototherapy procedures Cau...

Page 6: ...er is recommended In addition use of reflective foils may cause hazardous body temperatures Monitor the infant s skin temperature per your hospital policy during phototherapy to avoid fluctuations in...

Page 7: ...ous oxide or other anesthetic agents Warning Disconnect Electrical Power Always switch off the power and disconnect the power cord when cleaning the light Important Use of Nonstandard Components The u...

Page 8: ...neoBLUE mini System User Manual P N 051440D 5 Natus Manufacturer WEEE Directive Waste from Electrical and Electronic Equipment Directive Symbol Product should not be disposed of in normal waste...

Page 9: ...al safety by protecting the baby and the unit from incidental debris or fluids Arm The adjustable arm allows positioning of the light in multiple locations and angles The arm is adjustable in five loc...

Page 10: ...se check the intensity of the light using a spectrophotometer per your institution s procedures see Section 6 1 Checking the Light Intensity The light has been calibrated with the Ohmeda BiliBlanket M...

Page 11: ...inches 45 7 cm 22 25 W cm2 nm 48 square inches 309 7 cm Approximately 6 x 8 15 2 cm x 20 3 cm 4 3 Administering Phototherapy 1 Switch on power using the power switch Warning Operator Safety Sensitive...

Page 12: ...122 F Fan not working Verify that vents are not blocked If the lens surface has become hotter than 50 C 122 F verify that the fan is functioning correctly Baby becomes too warm Fan not working Verify...

Page 13: ...The light has been calibrated with the Ohmeda BiliBlanket Meter II to deliver a peak intensity of 30 35 W cm2 nm at a 12 inch 30 5 cm distance from the baby Please refer to section 9 of this manual or...

Page 14: ...eter The intensity of the light is set at the factory to 30 35 W cm2 nm at a 12 inch 30 5 cm distance from the light enclosure This measurement is taken at the central area of the effective surface ar...

Page 15: ...12 in over 6 hrs 18 F 10 C warmer than ambient at 30 5 cm over 6 hrs Electrical Mains 85 264 VAC 47 to 63 Hz Maximum Rating 0 7A 100 240V 50 60 Hz Safety Leakage current 100 A Audible Noise 60 dB Dim...

Page 16: ...tem can be placed closer to the infant yielding a higher intensity or higher dosage treatment To adjust the neoBLUE mini light to the desired intensity when changing the distance please refer to the s...