OLYMPIC SMART SCALE
®
Model 60/65
Page 3-13
Section 3: Service
Testing the Loadcell
1. With an empty tray on the scale, enter counts mode
(see below). If the displayed counts are outside of
500–5000, proceed to step 2.
2. Place a 7.5-kg weight on the tray and enter counts
mode. Adjust the VR4 until the counts shown are
32768 ±10 counts (see
Adjusting the Zero Offset
on
page 3-17).
If step 2 is not possible, return the scale to Natus
Medical for loadcell replacement.
Weight Lock Inhibit Mode
Pressing
gm/lb
twice enters Weight Lock Inhibit Mode; in
this mode, weights cannot lock on the display.
NOTE: Press off to exit test mode at any time.
Counts Mode
Simultaneously press
Reweigh
and power on the scale.
The raw AD counts show on the display. Press
Zero
to
momentarily show the calibration coefficient.
F
G
H
Summary of Contents for 60/65
Page 2: ......
Page 4: ......
Page 41: ...Page 3 22 OLYMPIC SMART SCALE Model 60 65 Page 3 22...
Page 42: ...Page 3 23 OLYMPIC SMART SCALE Model 60 65 Page 3 23...
Page 43: ...Page 3 24 OLYMPIC SMART SCALE Model 60 65 Page 3 24...
Page 44: ...Page 3 25 OLYMPIC SMART SCALE Model 60 65 Page 3 25...
Page 45: ...Page 3 26 OLYMPIC SMART SCALE Model 60 65 Page 3 26...
Page 46: ...Page 3 27 OLYMPIC SMART SCALE Model 60 65 Page 3 27...
Page 47: ...Page 3 28 OLYMPIC SMART SCALE Model 60 65 Page 3 28...
Page 48: ...Page 3 29 OLYMPIC SMART SCALE Model 60 65 Page 3 29...
Page 50: ......
Page 52: ......
Page 89: ...Page 3 22 OLYMPIC SMART SCALE Mod le 60 65 Page 3 22...
Page 90: ...Page 3 23 OLYMPIC SMART SCALE Mod le 60 65 Page 3 23...
Page 91: ...Page 3 24 OLYMPIC SMART SCALE Mod le 60 65 Page 3 24...
Page 92: ...Page 3 25 OLYMPIC SMART SCALE Mod le 60 65 Page 3 25...
Page 93: ...Page 3 26 OLYMPIC SMART SCALE Mod le 60 65 Page 3 26...
Page 94: ...Page 3 27 OLYMPIC SMART SCALE Mod le 60 65 Page 3 27...
Page 95: ...Page 3 28 OLYMPIC SMART SCALE Mod le 60 65 Page 3 28...
Page 96: ...Page 3 29 OLYMPIC SMART SCALE Mod le 60 65 Page 3 29...