background image

 

B. 

Place standard specimen slide (cover slip up) on top of stage surface.Swing moveable finger on slide 
holder outward.  Place specimen slide against fixed side of slide holder.  Slowly release moveable finger 
until it makes contact with specimen slide. 

 

C.  Rotate coarse focusing controls until specimen comes into focus. 

 

D.  Adjust fine focus controls until specimen is in sharp focus. 

 

E. 

Adjust diopter for difference in eyesight. 
 

a. 

Using right eye, peer into the right eyepiece tube.  Adjust sharpness of image by utilizing fine focus 
controls. 

 

b. 

Using left eye, peer into the left eyepiece tube.  Adjust sharpness of image by turning diopter 
adjustment located on left eyepiece tube. 

 

F. 

Adjusting the aperture (opening) of iris diaphragm. 

 

Iris diaphragm should not be used to control the brightness of illumination, use light intensity control 
knob to adjust light level. Iris diaphragms are designed to help achieve high resolution of specimen and 
provide contrast in the image.  Smaller apertures will deliver higher contrast to image.  However, closing 
aperture too much will reduce resolution.  Experimentation is the best method of determining the correct 
opening of diaphragm.  Some suggested openings for iris diaphragm are: 

 

OBJECTIVE

   

 

DIAPHRAGM OPENING

 

 

4x 

 

 

1/8 open  

(MIN) 

 

10x 

 

 

1/8 to 1/4 open 

 

40x 

 

 

1/4 to 1/2 open 

 

100x 

 

 

1/2 to 3/4 open  (MAX) 

 

G.  Changing magnification. 

 

 
 
 
 
 
 

 

a. 

Rotate revolving nosepiece to position 10x objective into optical path. 
 

b. 

This microscope has been parfocalized, which allows changes from one objective to another while 
requiring only a slight adjustment of the fine focus controls. 

 

c. 

When changing to the 40x and 100x objective lens, care must be exercised in order to prevent 
damaging the front lens element and specimen slide. 

 

d. 

In order to obtain maximum resolution of the 100x oil immersion lens, it is necessary to apply 
immersion oil between the coverglass of slide and front lens of the objective. 

 

1) 

Use of a very small amount of immersion oil is required. Only the very tip of the lens 

should ever come in contact with the immersion oil.  Oil should not come in contact 
with the white sealant ring on the objective.  Excess use of immersion oil will ruin your 
objective and void your warranty. 
 

2)  All air bubbles must be removed from between lens and slide by rotating nosepiece back and 

forth. 

 

H.  When finished viewing, all parts that come in contact with oil must be cleaned.  Failure to do so could 

permanently damage the 100x oil immersion objective lens.  Use of Windex to clean immersion oil off 
lens surfaces is recommended. 

 

I. 

Coarse focus tension adjustment. 
 

Objective Specification Chart 

Objective 

N.A. 

Color Code Ring  Field of View 

Magnification 

Din 4X 

0.10 

Red 

4.5mm 

40X 

Din 10X 

0.25 

Yellow 

1.8mm 

100X 

Din 40X 

0.65 

Blue 

0.45mm 

400X 

Din 100X 

1.25 

White 

0.18mm 

1000X 

Summary of Contents for DC5-169 Series

Page 1: ...nstruments Inc 6508 Tri County Parkway Schertz Texas 78154 Phone 210 590 9010 Fax 210 590 1104 INSTRUCTIONS FOR DC5 169 Series DIGITAL COMPOUND BIOLOGICAL MICROSCOPE Copyright 11 4 2016 National Optic...

Page 2: ...Mark on side of eyepiece tube for indexing diopter reading Light intensity control knob Base Illuminator condenser Knobs controlling X and Y movement of mechanical stage Fine focus knob Coarse focus k...

Page 3: ...ntainer in case the microscope needs transporting to another location or shipped for repairs Components are packed within the container as indicated below A Microscope body with 1 25 N A Abbe condense...

Page 4: ...ted into binocular eyepieces B Rod at mid position pull or push rod until you feel a gentle click stop 100 of microscope image is directed to built in camera C Rod pulled to fully extended position 30...

Page 5: ...1 2 open 100x 1 2 to 3 4 open MAX G Changing magnification a Rotate revolving nosepiece to position 10x objective into optical path b This microscope has been parfocalized which allows changes from o...

Page 6: ...rument on its side taking care to avoid damage to the specimen slide holder located on top of mechanical stage b Remove 5ea cross head screws holding base plate c Carefully lay the base plate flat on...

Page 7: ...le contents containing the full help menu Motic Live Imaging Module PC Full Help Menu The full Live Imaging Module manual is accessible within the live Imaging main page To begin open the Motic Images...

Page 8: ...ging Module help file containing the full help menu MOTIC IMAGES SOFTWARE MAC Full Help Menu The full software manual for Motic Images is accessible within the software s main page To begin open the M...

Page 9: ...re will allow you to view capture annotate and save your images For further assistance in using the Motic Images 3 0 software please refer to the Motic Help files These files will help explain the fun...

Page 10: ...ccess the Motic Images Live Imaging help menu click on Help located at the top left hand side of the screen Once the Help window open you will find the help guide within MODEL 926 PHASE CONTRAST SET I...

Page 11: ...ing knob to move microscope stage platform to its highest position 2 Remove objective dust caps from the revolving nosepiece 3 Screw objective lenses into nosepiece making certain to mount them in con...

Page 12: ...turret condenser until it is centered to the larger dark annulus located in the objective Both rings must be concentric to each other to achieve maximum performance Make sure that the knurled head cen...

Page 13: ...amera adapter Dirt on the protection filter of the camera sensor For Eyepieces with reticules Dirt on the outer or inner reticle glass Objectives are the optical component of the microscope that requi...

Page 14: ...e cloth moistened with a low graduation of alcohol If you face any problems related to the maintenance of your microscope please contact us Our technicians will gladly help you solve your maintenance...

Page 15: ...ave qualified service technician repair outlet IMAGE QUALITY PROBLEM REASON FOR PROBLEM SOLUTION No image Nosepiece not indexed properly Light too bright Move revolving nosepiece until objective lens...

Page 16: ...st objective lens 1 25 N A 704 160P DIN 4X Plan objective lens 0 10 N A 710 160P DIN 10X Plan objective lens 025 N A 740 160P DIN 40X Plan objective lens 0 65 N A 799 160P DIN 100X Plan objective lens...

Reviews: