Tension adjustment knob is located between stand and coarse focus knob of microscope, on the 
right side. 

 

b. 

To tighten tension of coarse focus knobs, turn control in a counter-clockwise direction. It is 
advisable to leave controls as loose as possible, tightening only enough to keep stage from drifting 
down and out of focus.  To loosen tension, turn control in clockwise direction. 

 

A. 

ELECTRICAL MAINTENANCE 
 

WARNING: FOR YOUR SAFETY, TURN SWITCH OFF AND REMOVE PLUG FROM POWER SOURCE 
OUTLET BEFORE MAINTAINING YOUR MICROSCOPE.

 

1. 

Replacement of lamp. 

a. 

Carefully lay instrument on its side, taking care to avoid damage to the specimen slide holder 
located on top of mechanical stage. 

b. 

Remove 5ea cross head screws, holding base plate. 

c. 

Carefully lay the base plate flat on the table. 

d. 

Disconnect the LED cable from the plug socket on the LED circuit card. 

 

 

 

e. 

Your microscope requires a 3W LED bulb assembly 

– Part # 800-500 

f. 

The LED bulb is located within the lamp housing.  It is held in place with screw on holder.  Simply 
unscrew the ring holder to remove the bulb assembly. 
 

 

 

g. 

Assemble the unit in the reverse order.  
 

MOTIC IMAGES SOFTWARE 

– PC Full Help Menu 

 

The full software manual for Motic Images is accessible within the software’s main page.

 

To begin, open the Motic Images Software. 

At the top of main screen find the menu tab labeled Help: 

Summary of Contents for DC5-169 Series

Page 1: ...nstruments Inc 6508 Tri County Parkway Schertz Texas 78154 Phone 210 590 9010 Fax 210 590 1104 INSTRUCTIONS FOR DC5 169 Series DIGITAL COMPOUND BIOLOGICAL MICROSCOPE Copyright 11 4 2016 National Optic...

Page 2: ...Mark on side of eyepiece tube for indexing diopter reading Light intensity control knob Base Illuminator condenser Knobs controlling X and Y movement of mechanical stage Fine focus knob Coarse focus k...

Page 3: ...ntainer in case the microscope needs transporting to another location or shipped for repairs Components are packed within the container as indicated below A Microscope body with 1 25 N A Abbe condense...

Page 4: ...ted into binocular eyepieces B Rod at mid position pull or push rod until you feel a gentle click stop 100 of microscope image is directed to built in camera C Rod pulled to fully extended position 30...

Page 5: ...1 2 open 100x 1 2 to 3 4 open MAX G Changing magnification a Rotate revolving nosepiece to position 10x objective into optical path b This microscope has been parfocalized which allows changes from o...

Page 6: ...rument on its side taking care to avoid damage to the specimen slide holder located on top of mechanical stage b Remove 5ea cross head screws holding base plate c Carefully lay the base plate flat on...

Page 7: ...le contents containing the full help menu Motic Live Imaging Module PC Full Help Menu The full Live Imaging Module manual is accessible within the live Imaging main page To begin open the Motic Images...

Page 8: ...ging Module help file containing the full help menu MOTIC IMAGES SOFTWARE MAC Full Help Menu The full software manual for Motic Images is accessible within the software s main page To begin open the M...

Page 9: ...re will allow you to view capture annotate and save your images For further assistance in using the Motic Images 3 0 software please refer to the Motic Help files These files will help explain the fun...

Page 10: ...ccess the Motic Images Live Imaging help menu click on Help located at the top left hand side of the screen Once the Help window open you will find the help guide within MODEL 926 PHASE CONTRAST SET I...

Page 11: ...ing knob to move microscope stage platform to its highest position 2 Remove objective dust caps from the revolving nosepiece 3 Screw objective lenses into nosepiece making certain to mount them in con...

Page 12: ...turret condenser until it is centered to the larger dark annulus located in the objective Both rings must be concentric to each other to achieve maximum performance Make sure that the knurled head cen...

Page 13: ...amera adapter Dirt on the protection filter of the camera sensor For Eyepieces with reticules Dirt on the outer or inner reticle glass Objectives are the optical component of the microscope that requi...

Page 14: ...e cloth moistened with a low graduation of alcohol If you face any problems related to the maintenance of your microscope please contact us Our technicians will gladly help you solve your maintenance...

Page 15: ...ave qualified service technician repair outlet IMAGE QUALITY PROBLEM REASON FOR PROBLEM SOLUTION No image Nosepiece not indexed properly Light too bright Move revolving nosepiece until objective lens...

Page 16: ...st objective lens 1 25 N A 704 160P DIN 4X Plan objective lens 0 10 N A 710 160P DIN 10X Plan objective lens 025 N A 740 160P DIN 40X Plan objective lens 0 65 N A 799 160P DIN 100X Plan objective lens...

Reviews: