3.  To change magnification: 

 

a.  Rotate revolving nosepiece to position 10x objective into optical path.  Some slight adjustment 

of focusing knob may be required. 

 

b.  Rotate revolving nosepiece to position 40x objective into optical path.  Some slight adjustment 

of focusing knob may be required. 

 

Note: Take care when rotating 40x objective into place.  This is the longest lens and has a 
spring retractable mechanism which retracts slightly into its housing if the front of the lens 
comes in contact with the specimen slide while focusing the microscope.  This prevents 
damage to the lens or slide. 

 

 
MAINTENANCE 

 

WARNING: For your own safety, turn switch to OFF position and remove plug from power source 
before maintaining your microscope.  If the power cord is worn, cut or damaged in any way, have 
it replaced immediately to avoid shock or fire hazard. 

 
A.  OPTICAL MAINTENANCE 
 

1.  Do not attempt to disassemble any lens components.  Consult an expert technical service 

company when repairs not covered by these instructions are needed. 

 

2.  Prior to cleaning any lens surface, brush dirt and lint off lens surface with camel hairbrush or 

compressed air.  Use of air in a can, available at most computer stores, is good source of clean 
air. 

 

3.  Do not remove eyepiece lens or objective lenses from microscope.  Clean only the outer lens 

surface by breathing on lens to dampen surface, then wipe with lens paper or cotton swab.  Avoid 
wiping lens surface while dry, as lenses are scratched very easily. 

 
B.  MECHANICAL MAINTENANCE 
 

1.  Stage stop adjustment: Stage stop has been pre-adjusted at the factory, and should not require 

readjustment. 

 

2.  Metal parts: Use a clean, damp cloth to remove dust or dirt from metal part followed by a dry cloth. 

 
C.  ELECTRICAL MAINTENANCE  
 

1.  Replacement of bulb (National #800-001)  To open the illuminator field lens housing use 0.90mm 

“L” type hex key wrench supplied with your microscope to loosen hex screws on lens housing.  
Remove lens housing to expose LED “bulb”.  Remove bulb by grasping the plastic base of bulb 
and gently pulling straight up.  Insert new LED “bulb”, replace lens housing and tighten hex screw 
to secure lens housing in place.  

 
2.  Replacement of fuse (National #801-160)   The fuse is located at right rear side of microscope 

base.  To remove fuse from holder, insert a 6mm screwdriver blade into slot located in rear of 
fuse holder cap.  Slightly depress and rotate screwdriver ¼ turn in direction of arrow to release the 
fuse.  Pull cap and fuse out of fuse holder.  Insert proper 0.5 amp fuse into fuse cap.  Insert fuse 
assembly into fuse holder.  Using screwdriver rotate fuse cap assembly in opposite direction of 
arrow until guide slot engages, depress fuse cap and rotate ¼ turn to lock into fuse holder. 

Summary of Contents for 104-CLED

Page 1: ...ific Instruments Inc 6508 Tri County Parkway Schertz Texas 78154 Phone 210 590 9010 Fax 210 590 1104 INSTRUCTIONS FOR MODEL 104 CLED ELEMENTARY COMPOUND MICROSCOPE Copyright 4 12 05 National Optical Scientific Instrument Inc ...

Page 2: ... eyepiece Eyepiece tube Objective turret nosepiece Objective lens Stage Stage clips Disc diaphragm Illuminator housing Illuminator switch Arm Stage stop screw Stage stop locking nut Focusing knob Base AC Cord Fuse ...

Page 3: ...ormed by the objective lens The inclined eyepiece is equipped with a pointer that rotates as the eyepiece is rotated B OBJECTIVE LENS Lens closest to the specimen forms the first magnified image of the specimen C OBJECTIVE TURRET nosepiece Revolving turret designed to hold objective lenses permits changes of magnification by rotating different powered objective lenses into optical path D STAGE CLI...

Page 4: ...t appear in field of view move specimen slide slightly on stage until image appears in field of view I Turn the disc diaphragm observing that different apertures affect the sharpness of the image Turn diaphragm until sharpest possible image is obtained When turning the disc diaphragm you should hear a click as each aperture comes into proper position under the hole in center of stage If aperture i...

Page 5: ...r objective lenses from microscope Clean only the outer lens surface by breathing on lens to dampen surface then wipe with lens paper or cotton swab Avoid wiping lens surface while dry as lenses are scratched very easily B MECHANICAL MAINTENANCE 1 Stage stop adjustment Stage stop has been pre adjusted at the factory and should not require readjustment 2 Metal parts Use a clean damp cloth to remove...

Page 6: ...7mm thick cover slip No 1 cover slip Slide upside down Place slide on stage with cover slip facing up Poor resolution Image not sharp Objective lenses dirty Clean objective lenses Eyepiece lens dirty Clean eyepiece lenses Too much light Adjust disc diaphragm Spots in field of view Eyepiece lens dirty Clean eyepiece lenses Specimen slide dirty Clean slide OPTIONAL ACCESSORIES AND PARTS 800 001 Repl...

Reviews: