background image

6

|

ni.com

|

NI WSN-3226 Calibration Procedure

14. Repeat steps 1 through 13 for each channel on the NI WSN-3226.

Note

The limits in Table 5 are derived using the values in Table 7.

If the Resistance Accuracy verification procedure determines that the NI WSN-3226 is outside 
of the limits, refer to 

Where to Go for Support

 for assistance in returning the device to NI.

Re-Verification

Repeat the 

Verification

 section to determine the As-Left status of the device.

Accuracy Under Calibration Conditions

The values in the following tables are based on calibrated scaling coefficients, which are stored 
in the onboard EEPROM.

The following accuracy table is valid for calibration under the following conditions:

Ambient temperature of 23 °C ± 5 °C

Relative humidity between 10% and 90%

Table 5.  

NI WSN-3226 Resistance Test Limits

RTD/Resistance 

Range Setting

Test Point (

)

3-Year Ranges

Lower Limit (

)

Upper Limit (

)

400 

/Pt100

100 

99.893

100.107

390

389.777

390.223

4 k

/Pt1000

1000

999.26

1000.74

4020

4018.35

4021.65

100 k

4020

4011.39

4028.21

100000

99963

100037

Table 6.  

NI WSN-3226 Voltage Accuracy

Range

Gain (% of Reading)

Offset (mV)

10 V

0.04

2.70

Summary of Contents for WSN-3226

Page 1: ...Accuracy Under Calibration Conditions 6 Where to Go for Support 7 Software Calibrating the NI WSN 3226 requires the installation of LabVIEW 2010 or later and NI WSN 1 2 or later on the calibration sy...

Page 2: ...evice are secure The NI WSN 3226 must be powered by either an external 9 V to 30 V power supply or four AA 1 5 V batteries alkaline only LabVIEW Help LabVIEW programming concepts and reference informa...

Page 3: ...llowing performance verification procedures describe the sequence of operation and test points required to verify the NI WSN 3226 and assume that adequate traceable uncertainties are available for the...

Page 4: ...Table 3 Note The limits in Table 3 are derived using the values in Table 6 If the Voltage Accuracy verification procedure determines that the NI WSN 3226 is outside of the limits refer to Where to Go...

Page 5: ...orresponds to the appropriate RTD resistance range 8 Turn any connection compensation settings on the calibrator off Tip On the Fluke 5520A press the COMP softkey to turn compensation off 9 Create a V...

Page 6: ...racy Under Calibration Conditions The values in the following tables are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calib...

Page 7: ...Specifications online at ni com manuals Where to Go for Support The National Instruments website is your complete resource for technical support At ni com support you have access to everything from t...

Reviews: