Appendix A
Specifications
VXIpc 770/870B Series User Manual
A-2
ni.com
Operating Environment
Ambient temperature range ....................5
to 50
°
C (Tested in accordance
with IEC-60068-2-1 and
IEC-60068-2-2.)
Relative humidity range..........................10 to 90%, noncondensing
(Tested in accordance with
IEC-60068-2-56.)
Altitude ...................................................2,000 m (at 25
°
C ambient
temperature)
Pollution Degree .....................................2
For indoor or enclosed area use only.
Caution
Clean the VXIpc-770 with a soft nonmetallic brush. Make sure that the device is
completely dry and free from contaminants before returning it to service.
Storage Environment
Ambient temperature range ....................–20
to 70
°
C (Tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2.)
Relative humidity range..........................5 to 95%, noncondensing
(Tested in accordance with
IEC-60068-2-56.)
Shock and Vibration
Operational shock ...................................30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)
Random vibration
Operating .........................................5 to 500 Hz, 0.3 g
rms
(with solid-state hard drive)
Nonoperating ...................................5 to 500 Hz, 2.4 g
rms
(Tested in accordance with IEC-60068-2-64. Nonoperating test profile