Chapter 11
VirtualBench-Counter
Getting Started with VirtualBench
11-8
www.ni.com
8.
Set the filter for input lines. This feature works only with devices with
NI-TIO based counters.
a.
Select Edit»Settings on the front panel.
b.
Select the Filter tab, shown in Figure 11-5.
Figure 11-5. Filter Tab of VirtualBench-Counter Settings Dialog Box
c.
Choose the Filter setting you want to use.
d.
If you use the Custom Filter, you must choose a counter to
generate the filter clock. Also, you must specify the size pulses to
allow to pass.
e.
With the built-in or custom filters, pulses greater than or equal to
the specified value are guaranteed to pass. Pulses half the size
specified are guaranteed to be blocked. Pulses between these sizes
may pass, depending on their phase with the filter clock. Refer to
the user manual of your DAQ device for more details.
f.
Click OK and continue with your measurements.
Note
Refer to the VirtualBench-Counter Online Reference for more information about
configuring VirtualBench-Counter for your specific application.