Table 1.

 LED Status Indication

LED Color

Status

 

 

Green

The NI RF power meter is connected to a computer or a chassis and is
functioning properly.

Yellow

An error has occurred or the firmware is being updated.

Off

The NI RF power meter is not initialized.

Front and Back Panel

Figure 2.

 NI 5683 Front Panel

1

1. Type-N Connector

Figure 3.

 NI 5683 Back Panel

3

1

2

4

1. Status LED
2. External Trigger In Terminal

3. USB 2.0 A to Mini-B Port
4. External Trigger Out Terminal

Note

 

External Trigger Out capability is currently unsupported.

Connector Care and Handling

The NI 5683 is a high-quality precision laboratory device and should be carefully handled.
Follow the precautions listed in this section when handling or connecting devices. Complying
with these precautions guarantees longer component life and lower equipment downtime
because of connector or device failure.

6

  | 

ni.com

  | 

NI 5683 Getting Started Guide

Summary of Contents for USB-5683

Page 1: ...3 if you use it in a manner not described in this document Contents Electromagnetic Compatibility Guidelines 1 Verifying the System Requirements 2 Unpacking the Kit 2 Verifying the Kit Contents 3 Prep...

Page 2: ...ft Caution To ensure the specified EMC performance the length of the USB cable must be no longer than 30 m 100 ft Verifying the System Requirements To use the NI 5683 your system must meet certain re...

Page 3: ...ontents you need the following items 1 N m torque wrench Desktop computer laptop or chassis with one available USB 2 0 or USB 1 1 compliant slot Preparing the Environment Ensure that the environment y...

Page 4: ...see access and security messages during installation Accept the prompts to complete the installation 4 When the installer completes select Restart in the dialog box that prompts you to restart shut d...

Page 5: ...ay need to use the included Type N to SMA adapter depending on the RF source connection Caution Always refer to the specifications document before connecting signals Failure to observe the specified m...

Page 6: ...l 3 1 2 4 1 Status LED 2 External Trigger In Terminal 3 USB 2 0 A to Mini B Port 4 External Trigger Out Terminal Note External Trigger Out capability is currently unsupported Connector Care and Handli...

Page 7: ...ermanently damages the internal components Before making connections verify the maximum input levels labeled below the ports on the module front panel Avoid Mechanical Shock Precision connectors are d...

Page 8: ...e device is still not listed power off the system ensure the device is correctly installed and restart 3 Record the device identifier MAX assigns to the hardware Use this identifier when programming t...

Page 9: ...n modify an NI 568x C example to create an application with Microsoft Visual C C and all required include and library files are automatically added to the project C or LabWindows CVI Available at Prog...

Page 10: ...pport ni com support RF Solutions ni com rf Updates ni com updates Services ni com services Getting Started with LabVIEW Getting Started with LabWindows CVI Located using the NI Example Finder Located...

Page 11: ...oC for your product by visiting ni com certification If your product supports calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corpora...

Page 12: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance p...