Derating Modules that Share the HVAB

Caution

As a safety precaution, modules that share a common HVAB 

must

 be derated to 

their lowest common voltage rating, even if they do not connect signals to the backplane. 

The following examples demonstrate how to derate switch modules:

Example 1

NI SCXI-1000 chassis
NI SCXI-1359 2-slot backplane adapter behind Slots 3 and 4
NI SCXI-1127 (300 V, CAT II) in Slot 4
NI SCXI-1129 (150 V, CAT I) in Slot 3

Both modules share the HVAB. Therefore, the NI SCXI-1127 must not 
exceed the rating of the NI SCXI-1129 (150 V, CAT I).

Example 2

NI SCXI-1000 chassis
NI SCXI-1359 2-slot backplane adapter behind Slots 3 and 4
NI SCXI-1127 (300 V, CAT II) in Slot 4
NI SCXI-1129 (150 V, CAT I) in Slot 2

The NI SCXI-1127 does not share the HVAB with the NI SCXI-1129. 
Therefore, the NI SCXI-1127 can operate at the full rating (300 V, CAT II).

Example 3

NI SCXI-1000 chassis
NI SCXI-1359 2-slot backplane adapter behind Slots 3 and 4
NI SCXI-1127 (300 V, CAT II) in Slot 4
NI SCXI-1190 (30 V, CAT I) in Slot 3

The NI SCXI-1190 does not have a HVAB connector. Therefore, the 
NI SCXI-1127 can operate at the full rating (300 V, CAT II).

Environment

The NI 1357/1358/1359 kit is intended for indoor use only.

Operating temperature............................ 0 

°

C to 50 

°

C

Storage temperature ............................... –20 

°

C to 70 

°

C

Relative humidity ................................... 10% to 90%, noncondensing

Measurement Category .......................... II

Summary of Contents for SCXI USB-1357

Page 1: ...USB 1357...

Page 2: ...ge signals among switches Introduction Note Both SCXI and USB kits connect to an SCXI switch with a 10 pin rear connector Refer to the NI Switches Getting Started Guide to determine the switches and s...

Page 3: ...I DAQ 7 4 or later before you install the NI 1357 1358 1359 kit Complete the following steps to install the NI 1357 1358 1359 kit 1 Power off and unplug the SCXI chassis Follow proper electrostatic di...

Page 4: ...the switch controller Refer to Figures 2 and 3 to verify the connector positions on the NI 1359 Figure 2 2 Slot HVAB Backplane Adapter Connectors Figure 3 8 Slot HVAB Backplane Adapter Connectors 1 HV...

Page 5: ...ts Connect one end of the SH9MD AUX cable to the NI SCXI 1359 cable adapter and the other end to the NI 4021 4060 407x switch controller as shown in Figure 4 Figure 4 Cabling an NI 4021 4060 407x Swit...

Page 6: ...configure the switch module Multichassis Configuration Complete the following steps to create a multichassis switch configuration 1 Complete the Installing the Kit section steps 1 through 7 for each c...

Page 7: ...of the HVAB backplane adapter and the other end to the NI 4021 4060 407x switch controller b Using another SH9MD AUX cable connect the AUX OUT of the HVAB backplane adapter on the first chassis to th...

Page 8: ...ure SCXI switch modules cabled to an NI SCXI 1357 1358 1359 kit proceed directly to step 4 to create a new SCXI chassis If you are using NI DAQmx to configure SCXI switch modules cabled to an NI USB 1...

Page 9: ...MAX detects only one communicating device this device is selected as the chassis communicator by default and the listbox is dimmed b SCXI kits only Select the module slot connected to the chassis com...

Page 10: ...it the default name click the Device Identifier textbox and enter a unique alphanumeric ID for the switch module c If you are using a connected accessory such as a terminal block select it from the Ac...

Page 11: ...witch module information A message at the top of the SCXI Chassis Configuration dialog box displays the configuration status If a warning appears in the message box you can save the configuration If a...

Page 12: ...NI SCXI 1129 150 V CAT I Example 2 NI SCXI 1000 chassis NI SCXI 1359 2 slot backplane adapter behind Slots 3 and 4 NI SCXI 1127 300 V CAT II in Slot 4 NI SCXI 1129 150 V CAT I in Slot 2 The NI SCXI 11...

Page 13: ...1 UL 61010 1 CAN CSA C22 2 No 61010 1 Note For UL and other safety certifications refer to the product label or visit ni com certification search by model number or product line and click the appropr...

Reviews: