SCXI-1520 Calibration Procedure
18
ni.com
Specification Tables
Table 2 contains the specifications you should use when verifying and
calibrating the gain and offset of the SCXI-1520. If the module has been
calibrated within the last year, the test point value should fall between the
and
values.
Table 2.
SCXI-1520 Gain and Offset Calibration Table
Test Point (V)
Gain
Lower Limit (V)
Upper Limit (V)
Index
4.900000
1
4.891100
4.908900
0
0.000000
1
–0.004000
0.004000
0
–4.900000
1
–4.908900
–4.891100
0
4.260870
1.15
4.253130
4.268609
1
0.000000
1.15
–0.003478
0.003478
1
–4.260870
1.15
–4.268609
–4.253130
1
3.769230
1.3
3.762385
3.776077
2
0.000000
1.3
–0.003077
0.003077
2
–3.769230
1.3
–3.776077
–3.762385
2
3.266670
1.5
3.260733
3.272600
3
0.000000
1.5
–0.002667
0.002667
3
–3.266670
1.5
–3.272600
–3.260733
3
2.722220
1.8
2.717278
2.727167
4
0.000000
1.8
–0.002222
0.002222
4
–2.722220
1.8
–2.727167
–2.717278
4
2.450000
2
2.445550
2.454450
5
0.000000
2
–0.002000
0.002000
5
–2.450000
2
–2.454450
–2.445550
5
2.227270
2.2
2.223227
2.231318
6
0.000000
2.2
–0.001818
0.001818
6
–2.227270
2.2
–2.231318
–2.223227
6
2.041670
2.4
2.037958
2.045375
7
0.000000
2.4
–0.001667
0.001667
7