SCXI-1303 Terminal Block Calibration Procedure

7.

Compare T

act

 to T

meas

.

If (T

meas

 

 0.5 

°

C) 

 T

act

 

 (T

meas

 

+

 0.5 

°

C), the performance of 

the terminal block temperature sensor has been verified.

If T

act 

<

 (T

meas

 

 0.5 

°

C),

 

the terminal block temperature sensor is 

nonfunctional. Do not substitute parts or modify equipment. 
Return the terminal block to National Instruments to ensure that 
the safety features are not compromised.

If T

act

 

>

 (T

meas

 

+

 0.5 

°

C), the terminal block temperature sensor is 

nonfunctional. Do not substitute parts or modify equipment. 
Return the terminal block to National Instruments to ensure that 
the safety features are not compromised.

8.

Return jumper W1 to its original position.

You have completed verifying the performance of the temperature sensor 
of your SCXI-1303 terminal block.