SCXI-1303 Terminal Block Calibration Procedure

4

ni.com

2.

Connect a calibrated DMM to the temperature-sensor output of the 
terminal block.

a.

Refer to Figure 2 to locate jumper W1 on your terminal block and 
verify that MTEMP is jumpered.

b.

Strip 0.5 inches of insulation from one end of a 22 AWG solid 
wire. Insert the stripped end of the wire into terminal A3 on the 
96-pin female DIN connector on the rear of the terminal block. 
Attach the other end of this wire to the positive input terminal of 
the calibrated DMM.

c.

Connect the negative input terminal of the calibrated DMM to the 
negative terminal of the +5 VDC power supply.

3.

Place the terminal block in a temperature-controlled environment 
where the temperature is between 15 and 35

°

C.

4.

When the terminal block temperature equilibrates with its 
surroundings, measure the temperature sensor output V

meas

 using 

a calibrated DMM.

5.

Measure the actual temperature T

act

 in the temperature-controlled 

environment using a calibrated thermometer.

6.

Convert V

meas

 (in volts) to measured temperature T

meas

 (in degrees

 

Celsius) by performing the following calculations:

a.

Calculate

b.

Calculate

c.

Calculate

where T

meas

 is in 

°

C

a = 1.295361 × 10

–3

b = 2.343159 × 10

–4

c = 1.018703 × 10

–7

x

2.5

V

me as

5000

---------------------------

=

y

V

m eas

x

-------------

ln

=

T

meas

1

a

y b

cy

2

+

(

)

+

----------------------------------

273.15

=