SCC-TC Series Thermocouple Input Modules User Guide
4
ni.com
Device Specific Information
Note
For general SCC module installation and signal connection information, and
information about the SC-2350 carrier, refer to the
SCC Quick Start Guide
, available
for download at
ni.com/manuals
.
Installing the Module
Caution
Refer to the
Read Me First: Safety and Radio-Frequency Interference
document
before removing equipment covers or connecting/disconnecting any signal wires.
You can plug the SCC-TC0
X
into any analog input socket on the SC-2345.
The socket you select determines which E Series DAQ device channels
receive the SCC-TC0
X
signals.
For single-stage input conditioning, plug the SCC-TC0
X
into any socket
J(
X
+1), where
X
is 0 to 7, and connect the input signals to the module as
described in the
If you use the SCC-TC0
X
in a dual-stage configuration, the SCC-TC0
X
must be the first-stage module. Plug it into any socket J(
X
+9) and plug the
second-stage SCC into socket J(
X
+1), where
X
is 0 to 7. Connect the input
signals to SCC-TC0
X
as described in the
section. The SC-2345 connects the output signals of the first-stage SCC to
the inputs of the second-stage SCC. An example of dual-stage conditioning
in an SCC-TC0
X
followed by an SCC-LP01 lowpass filter module.
Sockets J9 to J16 also are available for digital input/output (DIO)
conditioning or control. Refer to the
SC-2345 User Manual
for more
information on configuring, connecting, and installing SCC modules
Note
SCC-TC0
X
dual-stage configuration is only available in NI-DAQ 7.1 or later.