SCC-TC Series Thermocouple Input Modules User Guide

10

ni.com

Calibrating System Offsets

To calibrate the offset voltage of the SCC-TC0

X

 in the system using the 

E Series DAQ device, complete the following steps. The E Series DAQ 
device must be on for 30 minutes before you calibrate the offset voltage.

1.

Select the desired channel and gain on the E Series DAQ device.

2.

Short the inputs on the SCC-TC0

X

 screw terminals or miniplug.

3.

While acquiring data on the selected channel, use a screwdriver to 
adjust the potentiometer protruding through the top of the module until 
you read 0 VDC.

Calibrating the SCC-TC0

X

 Using a Thermocouple 

Calibrator

To calibrate the system using a thermocouple calibrator, complete the 
following steps:

1.

Make sure that the thermocouple connections at both the SCC and the 
thermocouple calibrator are at the same temperature.

2.

Connect the thermocouple calibrator to the SCC-TC0

X

. For best 

results, use thermocouple wire of the same length and type that you 
use in the thermocouple.

3.

Set the thermocouple calibrator to the required calibration 
temperature.

4.

Measure the calibrator voltage using the SCC-TC0

X

.

5.

Convert the measured calibrator voltage to a temperature measurement 
as described in the 

Scaling Voltage Measurements

 section.

6.

Adjust the potentiometer on the top of the SCC-TC0

X

 so that the 

measured temperature is equal to the calibration temperature.

Summary of Contents for SCC-TC Series

Page 1: ...bol leads you through nested menu items and dialog box options to a final action The sequence File Page Setup Options directs you to pull down the File menu select the Page Setup item and select Optio...

Page 2: ...a placeholder for a word or value that you must supply monospace Text in this font denotes text or characters that you should enter from the keyboard sections of code programming examples and syntax e...

Page 3: ...put Modules User Guide Read Me First Safety and Radio Frequency Interference SC 2345 2350 User Manual available at ni com SCC Quick Start Guide available at ni com SC 2345 Quick Reference Label 68 pin...

Page 4: ...ocket J X 1 where X is 0 to 7 and connect the input signals to the module as described in the Connecting the Input Signals section If you use the SCC TC0X in a dual stage configuration the SCC TC0X mu...

Page 5: ...odules function identically To use an SCC TC01 plug the male thermocouple connector into the module The SCC TC02 accepts up to three signals TC TC and GND TC is the positive thermocouple lead and TC i...

Page 6: ...Refer to Figure 1 for SCC TC0X signal connections Figure 1 SCC TC0X Signal Connections For more information about how to configure the SCC TC module using NI DAQmx refer to the SCC Quick Start Guide...

Page 7: ...IES is E Series DAQ device voltage This step provides proper scaling for the thermocouple amplifier in the SCC TC0X 2 Measure the reference junction cold junction temperature a Read the thermistor vol...

Page 8: ...lts of common mode voltage at the input of the SCC and current leakage into the signal leads The 10 M bias resistor in the SCC TC0X causes this error to be negligible With the 10 M bias resistor conne...

Page 9: ...he temperature in kelvin where a 1 295361 10 3 b 2 343159 10 4 c 1 018703 10 7 RT is the resistance of the thermistor in ohms where VTEMPOUT is the output voltage of the cold junction sensor where T F...

Page 10: ...ing the SCC TC0X Using a Thermocouple Calibrator To calibrate the system using a thermocouple calibrator complete the following steps 1 Make sure that the thermocouple connections at both the SCC and...

Page 11: ...dance 10 M powered on 10 k powered off or overload Bandwidth 2 Hz Amplifier Characteristics Open thermocouple detection current 250 nA max1 assuming floating thermocouple Common mode rejection ratio 1...

Page 12: ...V and reference junction measurement accuracy of 0 5 C Assumes averaging Non averaged single point reading has an additional uncertainty up to 0 1 C for J type thermocouple Thermocouple Type Temperatu...

Page 13: ...on Pull up resistor 10 M Bias resistor 10 M Maximum field wire gauge 16 AWG Power Requirement Analog power 60 mW max 15 V 2 mA max 15 V 2 mA max Digital power 5 V 0 0 mW Physical TC01 dimensions 7 8 c...

Page 14: ...cations refer to the product label or visit ni com hardref nsf search by model number or product line and click the appropriate link in the Certification column Electromagnetic Compatibility Emissions...

Page 15: ...2 shows the I O connector pins on the bottom of the module Figure 2 SCC Module Bottom View Table 1 lists the signal connection corresponding to each pin AI X and AI X 8 are the analog input signal cha...

Page 16: ...patents covering National Instruments products refer to the appropriate location Help Patents in your software the patents txt file on your CD or ni com patents 2002 2004 National Instruments Corp Al...

Reviews: