![National Instruments SCC-TC Series User Manual Download Page 13](http://html1.mh-extra.com/html/national-instruments/scc-tc-series/scc-tc-series_user-manual_3733614013.webp)
©
National Instruments Corporation
13
SCC-TC Series Thermocouple Input Modules User Guide
Note
The accuracy specification includes the combined effects of the temperature sensor
accuracy and the temperature difference between the temperature sensor and any
thermocouple connector. The temperature sensor accuracy includes component tolerances,
temperature drifts, and self-heating effects. It does not include measurement device errors.
Open Thermocouple Detection
Pull-up resistor ....................................... 10 M
Ω
Bias resistor............................................ 10 M
Ω
Maximum field wire gauge .................... 16 AWG
Power Requirement
Analog power ......................................... 60 mW max
+
15 V .............................................. 2 mA max
–15 V............................................... 2 mA max
Digital power (
+
5 V).............................. 0.0 mW
Physical
TC01 dimensions ................................... 7.8 cm
×
2.92 cm
×
1.85 cm
(3.1 in.
×
1.15 in.
×
0.73 in.)
TC02 dimensions ................................... 8.89 cm
×
2.92 cm
×
1.85 cm
(3.50 in.
×
1.15 in.
×
0.73 in.)
I/O .......................................................... One 20-pin right angle male
connector, 3-pin screw terminal
system (SCC-TC02);
two-prong mini-connector
(SCC-TC01)
Screw terminal field-wiring
gauge (SCC-TC02) ................................ 28 to 16 AWG
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the
common-mode voltage.
Each input must remain within
±
12 V of chassis ground.
Installation Category I