background image

Safety Voltages

Connect only voltages that are below these limits.

V terminal to C terminal

30 VDC maximum, Measurement Category I

Measurement Category I is for measurements performed on circuits not directly connected to
the electrical distribution system referred to as 

MAINS

 voltage. MAINS is a hazardous live

electrical supply system that powers equipment. This category is for measurements of voltages
from specially protected secondary circuits. Such voltage measurements include signal levels,
special equipment, limited-energy parts of equipment, circuits powered by regulated low-
voltage sources, and electronics.

Caution

 

Do not connect the sbRIO-9607 to signals or use for measurements within

Measurement Categories II, III, or IV.

Preparing the Environment

Ensure that the environment in which you are using the sbRIO-9607 meets the following
specifications.

Local ambient operating temperature near
device (IEC 60068-2-1, IEC 60068-2-2)

-40 °C to 85 °C

Maximum reported onboard sensor temperature

CPU/FPGA temperature

98 °C

Primary System temperature

85 °C

Secondary System temperature

85 °C

Note

 

Ensure that the local ambient, reported CPU/FPGA, and reported Primary

System temperatures do not exceed any of the maximum temperatures listed in this
document. For more information about how to access the onboard sensors, visit 

ni.com/info

 and enter the Info Code 

sbriosensors

.

Operating humidity (IEC 60068-2-78)

10% RH to 90% RH, noncondensing

Pollution Degree (IEC 60664)

2

Maximum altitude

5,000 m

Indoor use only.

Note

 

Refer to the device specifications on 

ni.com/manuals

 for complete

specifications.

2

  | 

ni.com

  | 

NI sbRIO-9607 Getting Started Guide

Summary of Contents for sbRIO-9607

Page 1: ...sbRIO 9607...

Page 2: ...Exercise caution when designing an enclosure for the sbRIO 9607 Auxiliary cooling may be necessary to keep the sbRIO 9607 within the specified operating temperature range Refer to the NI sbRIO 9607 S...

Page 3: ...ies II III or IV Preparing the Environment Ensure that the environment in which you are using the sbRIO 9607 meets the following specifications Local ambient operating temperature near device IEC 6006...

Page 4: ...ation from the kit Store the device in the antistatic package when the device is not in use Verifying the Kit Contents Verify that the following items are included in the sbRIO 9607 kit Figure 1 sbRIO...

Page 5: ...nfo and enter the Info Code softwareversion Connecting the sbRIO 9607 The sbRIO 9607 has the following components Figure 2 sbRIO 9607 Components 1 2 3 4 5 6 7 8 12 11 15 13 10 10 11 14 16 9 9 1 J9 Pow...

Page 6: ...til the connector latches into place 3 Turn on the power supply Powering On the NI sbRIO Device The NI sbRIO device runs a power on self test POST when you apply power to the device During the POST th...

Page 7: ...O 9607 is admin There is no default password for the sbRIO 9607 so you must leave the password field blank when logging in until you set a system password 1 Right click your system and select Web Conf...

Page 8: ...is complete Troubleshooting the sbRIO 9607 The sbRIO 9607 is Not Communicating with the Network Ensure that the Ethernet connections between the sbRIO 9607 and the host computer and the Ethernet conne...

Page 9: ...on for 5 s Run Mode Safe Mode Press and hold RESET button for 5 s Press and hold RESET button for 5 s Press and hold RESET button for 5 s Press and hold RESET button for 5 s Console Out enabled Networ...

Page 10: ...nto safe mode by pressing the reset button for longer than five seconds or by enabling safe mode in MAX Refer to the Measurement Automation Explorer MAX Help for information about safe mode Blinks fou...

Page 11: ...orldwide Support and Services The National Instruments website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application deve...

Page 12: ...ited States visit the Worldwide Offices section of ni com niglobal to access the branch office websites which provide up to date contact information support phone numbers email addresses and current e...

Page 13: ...d third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance pol...

Reviews: