Required current
150 mA
Safety Voltage
Connect only voltages that are below these limits.
Voltage
|Peaks| ≤20 V
Figure 3. SA1000X Maximum Input Voltage Versus Frequency. Derate voltage according to the
frequency of the signal you are measuring.
1 M
10 M
10
100
1
100 M
1 G
10 G
Frequency (Hz)
V
oltage
(V peak AC)
Maximum Input Voltage
Dynamic Measuring Range
Measurement Category
CAT O
Caution
Do not connect the SA1000X to signals or use for measurements
within Measurement Categories II, III, or IV.
Attention
Ne connectez pas le SA1000X à des signaux et ne l'utilisez pas
pour effectuer des mesures dans les catégories de mesure II, III ou IV.
Measurement Category I is for measurements performed on circuits not directly
connected to the electrical distribution system referred to as MAINS voltage. MAINS
is a hazardous live electrical supply system that powers equipment. This category is
for measurements of voltages from specially protected secondary circuits. Such
ni.com
10
SA1000X Safety, Environmental, and Regulatory Information