Installation
Mounting
This product is designed for mounting in an EIA-310 compliant 19 in. rack. It includes
adjustable rack ears for mounting in either a flush or recessed configuration. The user is
responsible for ensuring the mounting method provides adequate structural support for the
product and any attached cables. Ensure the product is securely mounted before applying power.
Ventilation
The user is responsible for ensuring the mounting location provides adequate ventilation to
dissipate heat generated during operation of the product. Some models include an air intake on
the front panel and an exhaust fan on the rear panel. Ensure no external equipment or cables
restrict the airflow through the intake or exhaust ports.
Chassis Ground
The rear of the chassis includes a redundant chassis ground screw and ground wire. If desired
for your installation location, connect the chassis ground wire to the rack cabinet using an
appropriate fastener.
Main CB Off
Ensure the main CB on the front of the product is in the OFF position before connecting mains
power.
Facility Power Source
The facility power source or upstream equipment feed for this product must include an
overcurrent protective device with a maximum rating as shown in the following table.
Table 5.
Required Facility or Equipment Overcurrent Protection
Model (Part Number)
Related
Continuous Current
Required
Mains Protection
RMX-10050 (785341-01, 785342-01)
16 A
20 A
RMX-10051 (785344-01)
16 A
20 A
RMX-10051 (785343-01)
16 A
20 A