Following the adjustment procedure automatically updates the calibration date
and temperature in the external calibration area of the onboard flash memory of
the NI PXIe-7865/7866 device.

If an adjustment is not required use the

nirio78xxCalibrationUpdateExtCalDate.vi

 to update the External

Calibration date.

Complete the following procedure to adjust the calibration constants performance of the NI
PXIe-7865/7866 device.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect your device to the calibrator.

3.

Initialize the external calibration process by calling

nirio78xxRCalibrationAdjustExtCal.vi

. Set the Action to 

Initialize

. This

action is password-protected and the default password is 

NI

. You do not need to input

values for Comments or Current Calibrator Value.

4.

Determine the value of the NI PXIe-7865/7866 device onboard reference by completing
the following steps.
a)

Call 

nirio78xxCalibrationAdjustExtCal.vi

. Set the Action to

CalibrateOnboardReference

 and pass the Current Calibrator Value of 0 V to

the VI for the first iteration. 

nirio78xxCalibrationAdjustExtCal.vi

 then

performs measurements and determines a new Requested Calibrator Value.

b)

Program the calibrator to output the Requested Calibrator Value. At this point, you
must wait to ensure that the calibrator output has settled completely.

c)

Repeat steps a and b until the Reference Calibration Done output becomes TRUE.
With each new iteration of these steps, input the Requested Calibrator Value returned
from the previous iteration.

5.

Adjust the NI PXIe-7865/7866 device external calibration constants by calling

nirio78xxCalibrationAdjustExtCal.vi

. Set the Action to

AdjustCalibration

.

6.

Save the calibration adjustment values to the onboard flash memory using

nirio78xxCalibrationAdjustExtCal.vi

. Set the Action to 

CommitAndClose

.

a)

If you would prefer to cancel the calibration, you call this VI and set the Action to

CancelandClose

. A comment input is provided for storing any comments related

to your calibration.

7.

Disconnect the calibrator from the device.

Reverification

Repeat the 

Verification

 section to determine the As-Left status of the device.

Note

 

If any test fails Reverification after performing an adjustment, verify that you

have met the Test Conditions before returning your device to NI. Refer to the

Worldwide Support and Services

 section for assistance in returning the device to NI.

8

  | 

ni.com

  | 

NI PXIe-7865 and NI PXIe-7866 Calibration Procedure

Summary of Contents for PXIe-7865

Page 1: ...Device Drivers January 2020 or later Documentation Consult the following documents for information about your device LabVIEW and driver software All documents are available on ni com manuals and help...

Page 2: ...ppm or less DMM NI 4070 Use a multiranging 6 digit DMM with an accuracy of 250 ppm Counter Agilent 53131A Use a counter accurate to 0 01 Low thermal copper EMF plug in cables Fluke 5440A 7002 Do not...

Page 3: ...ew vi lib LabVIEW Targets FPGA RIO R Series 78XXR 783XR Calibration Public Self Calibration Complete the following steps to self calibrate the device Note No signal connections are needed for self cal...

Page 4: ...to the calibrator as shown in the following figure Figure 1 Analog Input Verification Connections Calibrator HI LO AI1 AI0 2 If your calibrator has a guard connection connect that terminal to AI GND 3...

Page 5: ...0 Negative full scale 5 5 000820 4 999180 5 001170 4 998830 2 2 Positive full scale 2 1 999661 2 000339 1 999521 2 000479 0 0 0 000137 0 000137 0 000137 0 000137 Negative full scale 2 2 000339 1 99966...

Page 6: ...erification Test Limits table 6 Disconnect the DMM from the NI R Series device 7 Repeat steps 1 through 6 for all AO channels on the device Table 3 Analog Output Accuracy Verification Test Limits Rang...

Page 7: ...ies Counter Verification Test Limits Set Point MHz Lower Limit MHz Upper Limit MHz 5 0 4 999500 5 000500 5 Stop the generation by calling nirio78xxCalibrationStopCounter vi 6 Disconnect the counter fr...

Page 8: ...mines a new Requested Calibrator Value b Program the calibrator to output the Requested Calibrator Value At this point you must wait to ensure that the calibrator output has settled completely c Repea...

Page 9: ...ni com services for information about the services NI offers Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you receive importan...

Page 10: ...ments EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy...

Reviews: