Table 2. Analog Input Accuracy Verification Test Limits
Range (V)
Test Point
24-Hour Limits
1-Year Limits
Minimum Maximum Location
Value
(V)
Lower
Limit (V)
Upper
Limit (V)
Lower
Limit (V)
Upper
Limit (V)
-10
10
Positive
full-scale 10
9.998416
10.001584 9.997717
10.002283
0
0
-0.000639
0.000639
-0.000639
0.000639
Negative
full-scale -10
-10.001584 -9.998416 -10.002283 -9.997717
-5
5
Positive
full-scale 5
4.999180
5.000820
4.998830
5.001170
0
0
-0.000340
0.000340
-0.000340
0.000340
Negative
full-scale -5
-5.000820
-4.999180 -5.001170
-4.998830
-2
2
Positive
full-scale 2
1.999661
2.000339
1.999521
2.000479
0
0
-0.000137
0.000137
-0.000137
0.000137
Negative
full-scale -2
-2.000339
-1.999661 -2.0004790 -1.999521
-1
1
Positive
full-scale 1
0.999817
1.000183
0.999748
1.000252
0
0
-0.000074
0.000074
-0.000074
0.000074
Negative
full-scale -1
-1.000183
-0.999817 -1.000252
-0.999748
Analog Output Accuracy Verification
Complete the following procedure to determine the As-Found status of the NI
PXIe-7865/7866.
1.
Connect the NI R Series device to the DMM as shown in the following figure.
NI PXIe-7865 and NI PXIe-7866 Calibration Procedure
|
© National Instruments
|
5