background image

Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at 

ni.com/trademarks

 for

information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their
respective companies. For patents covering NI products/technology, refer to the appropriate location: 

Help

»

Patents in your

software, the 

patents.txt

 file on your media, or the National Instruments Patent Notice at 

ni.com/patents

. You can find

information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer
to the Export Compliance Information at 

ni.com/legal/export-compliance

 for the NI global trade compliance policy and how

to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS
TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S.
Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable
limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2020 National Instruments. All rights reserved.

378236A-01

April 9, 2020

Summary of Contents for PXIe-7865

Page 1: ...Device Drivers January 2020 or later Documentation Consult the following documents for information about your device LabVIEW and driver software All documents are available on ni com manuals and help...

Page 2: ...ppm or less DMM NI 4070 Use a multiranging 6 digit DMM with an accuracy of 250 ppm Counter Agilent 53131A Use a counter accurate to 0 01 Low thermal copper EMF plug in cables Fluke 5440A 7002 Do not...

Page 3: ...ew vi lib LabVIEW Targets FPGA RIO R Series 78XXR 783XR Calibration Public Self Calibration Complete the following steps to self calibrate the device Note No signal connections are needed for self cal...

Page 4: ...to the calibrator as shown in the following figure Figure 1 Analog Input Verification Connections Calibrator HI LO AI1 AI0 2 If your calibrator has a guard connection connect that terminal to AI GND 3...

Page 5: ...0 Negative full scale 5 5 000820 4 999180 5 001170 4 998830 2 2 Positive full scale 2 1 999661 2 000339 1 999521 2 000479 0 0 0 000137 0 000137 0 000137 0 000137 Negative full scale 2 2 000339 1 99966...

Page 6: ...erification Test Limits table 6 Disconnect the DMM from the NI R Series device 7 Repeat steps 1 through 6 for all AO channels on the device Table 3 Analog Output Accuracy Verification Test Limits Rang...

Page 7: ...ies Counter Verification Test Limits Set Point MHz Lower Limit MHz Upper Limit MHz 5 0 4 999500 5 000500 5 Stop the generation by calling nirio78xxCalibrationStopCounter vi 6 Disconnect the counter fr...

Page 8: ...mines a new Requested Calibrator Value b Program the calibrator to output the Requested Calibrator Value At this point you must wait to ensure that the calibrator output has settled completely c Repea...

Page 9: ...ni com services for information about the services NI offers Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you receive importan...

Page 10: ...ments EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy...