1.
Launch MAX.
2.
In the configuration tree, expand
Devices and Interfaces
to see the list of installed
devices.
Installed devices appear under the name of their associated chassis.
3.
Expand your
Chassis
tree item.
MAX lists all devices installed in the chassis. Your default device names may vary.
Note
If you do not see your device listed, press <F5> to refresh the list of
installed devices. If the device is still not listed, power off the system, ensure
the device is correctly installed, and restart.
4.
Record the device identifier MAX assigns to the hardware. Use this identifier when
programming the NI 5698.
5.
To use the NI 5698 with the NI 5668R vector signal analyzer, you must associate the
NI 5698 with the NI 5606 RF downconverter in MAX.
Note
This step is not required if you are using the NI 5698 as a stand-alone
device.
a)
Select the system device with the appropriate downconverter in the configuration
tree.
b)
In the Associated Devices section, use the
RF Conditioning
drop-down listbox to
select the NI 5698.
c)
Click
Save
in the MAX toolbar.
Note
Module associations may be lost when you move the modules to
different chassis slots.
6.
Self-test the device by selecting the device in the configuration tree and clicking
Self-Test
in the MAX toolbar.
The MAX self-test performs a basic verification of hardware resources.
Related Information
Refer to the NI RF Vector Signal Analyzers Help for more information about renaming
devices.
Troubleshooting