Figure 1.
Chassis Compatibility Symbols
NI PXIe-1062Q
1
2
3
4
5
1. PXI Express System Controller Slot
2. PXI Peripheral Slot
3. PXI Express Hybrid Peripheral Slot
4. PXI Express System Timing Slot
5. PXI Express Peripheral Slot
The NI 5698 module can be placed in PXI Express peripheral slots or PXI Express
Hybrid peripheral slots.
6.
Touch any metal part of the chassis to discharge static electricity.
7.
Ensure that the ejector handle is in the downward (unlatched) position.
8.
Place the module edges into the module guides at the top and bottom of the chassis. Slide
the device into the slot until it is fully inserted.
Figure 2.
Module Installation
2
3
NI PXIe-1075
1
1. Chassis
2. Hardware Module
3. Ejector Handle in Downward (Unlatched) Position
9.
Latch the module in place by pulling up on the ejector handle.
10. Secure the device front panel to the chassis using the front-panel mounting screws.
Note
Tightening the top and bottom mounting screws increases mechanical
stability and also electrically connects the front panel to the chassis, which can
improve the signal quality and electromagnetic performance.
11. Cover all empty slots with filler panels or use slot blockers to maximize cooling air flow.
NI PXIe-5698 Getting Started Guide
|
© National Instruments
|