NI
P
X
Ie
-56
94 Calibr
a
tion Pro
c
e
dure
3.5 mm (m)-to-
3.5 mm (m) adapter
Huber+Suhner
32_PC35-50-0-2/
199_NE
Verifying third-order intercept point
out-of-band
Adjusting IF gain for bypass path
Adjusting IF gain
Adjusting IF flatness
Frequency range: DC to 33 GHz
Impedance: 50
Ω
Return loss:
DC to 1.5 GHz......................
≥
35 dB
1.5 GHz to 6.0 GHz..............
≥
30 dB
6.0 GHz to 18.0 GHz............
≥
20 dB
3.5 mm (f)-to-
3.5 mm (f) adapter
Huber+Suhner
31_PC35-50-0-2/
199_N
Characterizing cable loss
Frequency range: DC to 18 GHz
Impedance: 50
Ω
Return loss:
DC to 1.5 GHz......................
≥
35 dB
1.5 GHz to 6.0 GHz..............
≥
30 dB
6.0 GHz to 18.0 GHz............
≥
20 dB
BNC (m)-to-
BNC (m) cable
(quantity: 3)
—
Verifying third-order intercept point
out-of-band
Adjusting IF gain for bypass path
Adjusting IF gain
Adjusting IF flatness
Length: 36 in.
Impedance: 50
Ω
Table 1.
Recommended Equipment for NI 5694 Calibration (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements