NI PXIe-5694 Calibration Procedure
|
© National Instruments
|
33
Adjusting IF Gain for Bypass Path
Complete the following procedure to measure and adjust the IF gain for bypass path
performance of the NI 5694 IF conditioning module.
1.
Zero and calibrate the power sensors using the built-in functions in the power meter.
2.
Ensure that the hardware is properly configured as described in
3.
Run the self-calibration procedure for the NI 5622 digitizer.
4.
Initialize an external calibration session for the NI 5694.
5.
Initialize an IF attenuation calibration step.
6.
Configure the NI 5694 as follows:
•
Signal conditioning: Bypassed
•
IF conditioning downconversion: Disabled
•
Step gain: Disabled
7.
Configure RF source 1 as follows:
•
Single frequency mode
•
Frequency: 107.5 MHz
•
Power level: 0 dBm
8.
Configure the power meter as follows:
•
Channel 1: Power sensor A
•
Trigger with settling delay
•
Sensor setting: 0.1%
9.
Configure the NI 5622 digitizer as follows:
•
Sampling rate: 585.9375 kS/s
•
Number of samples to acquire: 40,960
•
Reference source: PXI_Clk
•
Reference Clock rate: 100 MHz
•
DDC enabled: TRUE
•
Data processing mode: Complex
•
Frequency translation enables: TRUE
•
DDC center frequency: 107.5 MHz
•
Dither enabled: FALSE
•
Bandpass filter enabled: FALSE
10. Read the power meter channel A power.
11. Calculate the NI 5694
IF input power
using the following equation:
IF Input Power
=
Channel A Power
+
Splitter Tracking at RF Source Frequency