Reinstalling the DC Block
Figure 7. Installing the DC Block on the PXIe-5606 RF IN Connector
1
1
2
3
4
1. Screws
2. DC block support
3. DC block
4. PXIe-5606 RF IN Connector
1.
Attach the DC block to the RF IN connector on the PXIe-5606 downconverter module
with your hand.
2.
Tighten the DC block with a torque wrench set to a maximum torque of 1 N · m
(9 lb · in.).
Notice
Using more than the recommended amount of torque may damage the
RF IN connector.
3.
Place the DC block support around the DC block and align the mounting holes with the
holes on the PXIe-5606 as shown in the figure above.
4.
Insert two screws through the DC block support mounting holes into the holes on the
PXIe-5606 as shown in the figure above.
5.
Tighten the screws with a Torx screwdriver with a T8 bit to a torque of 1.65 N · m
(5 lb · in.).
16
|
ni.com
|
PXIe-5668 Getting Started Guide