Summary of Contents for PXIe-5667

Page 1: ...PXIe 1075...

Page 2: ...RF preselector module NI PXIe 5694 NI 5694 IF conditioning module Caution The protection provided by this product may be impaired if it is used in a manner not described in this document Hot Surface I...

Page 3: ...ications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules Caution To ensure the specified EMC performance operate...

Page 4: ...id SMA to SMA Cable x2 Labeled N Part Number 154048A 01 7 Medium Semi Rigid SMA to SMA Cable x3 Labeled J Part Number 151611A 01 8 Long Semi Rigid SMA to SMA Cable Labeled P Part Number 154049A 01 9 F...

Page 5: ...7 in meets the following specifications Operating ambient temperature IEC 60068 2 1 IEC 60068 2 2 0 C to 55 C Operating relative humidity IEC 60068 2 56 10 to 90 noncondensing Maximum altitude 2 000 m...

Page 6: ...d cables you must install the NI 5694 IF conditioning module immediately to the left of the NI 5603 RF downconverter module the NI 5622 IF digitizer module immediately to the left of the NI 5694 modul...

Page 7: ...he top and bottom of the chassis Slide the device into the slot until it is fully inserted Figure 2 Module Installation 2 3 NI PXIe 1075 1 1 Chassis 2 Hardware Module 3 Ejector Handle in Down Unlatche...

Page 8: ...GHz OUT ALL PORTS 50 REF OUT 100 MHz 1 5 V p p MAX ACCESS ACTIVE IFConditioningModule NI PXIe 5694 ESD SENSITIVE REF LO IN 10 MHz 215 MHz 10 dBm NOM 18 dBm MAX 12VDC MAX IF OUT 187 5 MHz 193 6 MHz 21...

Page 9: ...plete tightening all SMA connectors to 1 N m using an appropriate torque wrench not included or torque screwdriver and SMA driver bit Tighten only until the wrench clicks Caution Incorrect torque at S...

Page 10: ...IN connector ensure that there are no signal transients greater than the RF and DC specifications for the device that are being generated and sourced to the NI 5603 RF IN connector NI 5603 RF Downcon...

Page 11: ...t 20 dBm maximum at 10 MHz and at 0 dBm maximum at 10 MHz To properly configure the input attenuators set the reference level at or above the input signal level with the Reference Level property or th...

Page 12: ...minal and a maximum safe power level of 15 dBm The maximum safe voltage is 25 VDC LO2 OUT Output terminal for the LO2 4 GHz source In multichannel systems LO2 OUT exports the signal received at LO2 IN...

Page 13: ...programmed by NI RFSA ACTIVE Indicates the module state OFF The module is in a quiescent state AMBER The module is waiting for the Advance Trigger from the configuration list GREEN The module is trig...

Page 14: ...ntains four connectors and two LEDs Figure 5 NI 5622 IF Digitizer Module Front Panel ACCESS ACTIVE CLK IN CLK OUT PFI 1 ESD SENSITIVE TTL 6 3 Vp p MAX 2 Vp p NOM 20 dBm MAX 50 50 50 16 Bit IF Digitize...

Page 15: ...7 the IF IN connector has a 4 dBm full scale input level and a 6 dBm input level nominal The maximum input level is 20 dBm or damage can occur CLK IN Connects to the 100 MHz REF OUT connector on the N...

Page 16: ...RFSA RED The module has exceeded approved operating temperature and thermal shutdown has occurred ACTIVE Indicates the module state OFF The module is not armed triggered or acquiring a waveform AMBER...

Page 17: ...EDs Figure 6 NI 5653 Synthesizer LO Source Module Front Panel ESD SENSITIVE ACCESS ACTIVE LO3 800 MHz LO2 4 GHz LO1 3 2 GHz 8 3 GHz OUT ALL PORTS 50 NI PXIe 5653 Synthesizer REF IN 10 MHz 5 V p p MAX...

Page 18: ...ximum voltage of 2 Vpk pk REF OUT 10 MHz Routes a frequency reference signal from the synthesizer LO module onboard 10 MHz oven controlled crystal oscillator OCXO REF OUT 100 MHz Connects to the CLK I...

Page 19: ...is not generating a signal AMBER The module phased locked loops PLLs are attempting to lock GREEN The module is generating a signal applicable PLLs are locked RED The module has detected an error sta...

Page 20: ...NSITIVE Table 10 Device Front Panel Icon Definitions Refer to the user documentation for required maintenance measures to ensure user safety and or preserve the specified EMC performance The signal pi...

Page 21: ...ates the basic hardware status of the module OFF The module is not yet functional or the module has detected a problem with a power rail AMBER The module is being accessed Accessed means that the devi...

Page 22: ...VDC MAX REF OUT 10 MHz 10 dBm NOM 12 VDC MAX LO OUT 215 MHz 10 dBm NOM 12 VDC MAX Table 13 Device Front Panel Icon Definitions Refer to the user documentation for required maintenance measures to ensu...

Page 23: ...EF LO IN Connects to the REF OUT 10 MHz connector on the NI 5653 front panel when used as a reference signal to the internal 215 MHz PLL in downconversion mode Connect the REF LO IN connector to a 215...

Page 24: ...the module has exceeded approved operating temperature and thermal shutdown has occurred or that the module has detected a power supply failure If the power supply fails contact NI technical support C...

Page 25: ...ep for all modules in your NI 5667 system The MAX self test performs a basic verification of hardware resources Related Information Troubleshooting on page 28 If an issue persists after you complete a...

Page 26: ...er self calibration 5 C and or 90 days Performing a Device Self Calibration Using the NI RFSA SFP NI recommends you run the self calibration from the NI RFSA Soft Front Panel SFP You can also run a se...

Page 27: ...ming the NI 5667 You can acquire data interactively using the NI RFSA Soft Front Panel SFP or use NI RFSA to program your device in the ADE of your choice 26 ni com NI PXIe 5667 3 6 GHz Getting Starte...

Page 28: ...located in the NIDocDir NI RFSA examples directory where NIDocDir is one of the following directories Windows 8 7 Vista Users Public Public Documents National Instruments Windows XP Documents and Set...

Page 29: ...ignal generators and NI toolkit software including the Modulation Toolkit are also available online at ni com examples Troubleshooting If an issue persists after you complete a troubleshooting procedu...

Page 30: ...PXI controller verify that a National Instruments entry appears in the system device list Reinstall NI RFSA and the device if error conditions appear in the list If you are using an MXI controller ri...

Page 31: ...or visit ni com support What Should I Do if the NI RFSA Soft Front Panel Does Not Launch 1 Verify the front panel interconnections for your RF spectrum monitoring receiver 2 Verify the MAX device ass...

Page 32: ...ics Getting Started with LabWindows CVI Support ni com support This item is also installed with the driver software EXPLORE LEARN CREATE DISCOVER RF Solutions ni com rf Services ni com services NI Com...

Page 33: ...office websites which provide up to date contact information support phone numbers email addresses and current events Refer to the NI Trademarks and Logo Guidelines at ni com trademarks for informati...

Reviews: