7. Ensure that the ejector handle is in the downward (unlatched) position.
8. Place the module edges into the module guides at the top and bottom of the chassis. Slide
the module into the slot until it is fully inserted.
Figure 3.
Module Installation
2
3
NI PXIe-1075
1
1. Chassis
2. Hardware Module
3. Ejector Handle in Downward (Unlatched) Position
9. Latch the module in place by pulling up on the ejector handle.
10. Secure the module front panel to the chassis using the front-panel mounting screws.
Note
Tightening the top and bottom mounting screws increases mechanical
stability and also electrically connects the front panel to the chassis, which can
improve the signal quality and electromagnetic performance.
11. Cover all empty slots using EMC filler panels or fill using slot blockers to maximize
cooling air flow, depending on your application.
12. Power on the chassis.
Related Information
on page 6
Interconnecting the PXIe-5665 Modules
Complete the following steps to interconnect the PXIe-5665 front panel connectors using the
PXIe-5603 downconverter, as shown in Figure 3, or the PXIe-5605 downconverter, as shown
in Figure 4.
Caution
The signal pins of this product's input/output ports can be damaged if
subjected to ESD. To prevent damage, remove power from the product before
connecting cables and employ industry-standard ESD prevention measures during
installation, maintenance, and operation.