5. Power on the chassis.
6. Verify that the device appears in MAX.
7. Reset the device in MAX and perform a self-test.
What Should I Do if the PXIe-5665 Doesn't Appear in MAX?
1. In the MAX configuration tree, expand
Devices and Interfaces
.
2. Expand the
Chassis
tree to see the list of installed hardware, and press <F5> to refresh
the list.
3. If the module is still not listed, power off the system, ensure that all hardware is correctly
installed, and restart the system.
4. Navigate to the Device Manager.
Operating System
Description
Windows 10/8.1
Right-click the Start button, and select
Device Manager
.
Windows 7
Select
Start
»
Control Panel
»
Device Manager
.
5. Verify the PXIe-5665 appears in the Device Manager.
a) Under an NI entry, confirm that a PXIe-5665 entry appears.
Note
If you are using a PC with a device for PXI remote control system,
under
System Devices
, also confirm that no error conditions appear for the
PCI-to-PCI Bridge
.
b) If error conditions appear, reinstall NI-RFSA and the PXIe-5665.
What Should I Do if the PXIe-5665 Fails the Self-Test?
1. Restart the system.
2. Launch MAX, and perform the self-test again.
3. Power off the chassis.
4. Reinstall the failed module in a different slot.
5. Power on the chassis.
6. Perform the self-test again.
Configuration
What Should I Do if the Device Does Not Initialize?
Failure to initialize may indicate a problem with module interconnections or with MAX. If the
niRFSA Initialize VI or the
niRFSA_init
function returns an error and the PXIe-5665 fails
to initialize, complete the following steps:
1. Reconnect the PXIe-5665 hardware module front panel cables securely.
28
|
ni.com
|
PXIe-5665 Getting Started Guide