Caution
Apply external signals only while the NI 5632 is powered on. Applying
external signals while the device is powered off may cause damage.
•
Ensure you are properly grounded when manipulating cables or antennas connected to the
NI 5632 PORT 1 or PORT 2 connector.
•
If you are using noninsulated devices, such as a noninsulated RF antenna, ensure the
devices are maintained in a static-free environment.
•
If you are using an active device, such as a preamplifier or switch routed to the NI 5632
PORT 1 or PORT 2 connector, ensure that there are no signal transients greater than the
RF and DC specifications for the device that are being generated and sourced to the
NI 5632 PORT 1 or PORT 2 connector.
NI 5632 Front Panel
The NI 5632 contains nine connectors and two LEDs.
Figure 5.
NI PXIe-5632 Front Panel
NI PXIe-5632
300 kHz - 8.5 GHz Vector Network Analyzer
Dual Sources
PORT 1
SOURCE 1
IN
SOURCE 2
IN
LO OUT
PFI 0
TTL
SOURCE 2
OUT
ACCESS
ACTIVE
+27 dBm MAX
50 VDC
+27 dBm MAX
0 VDC
+27 dBm MAX
0 VDC
+27 dBm MAX
0 VDC
+20 dBm MAX
50 VDC
+27 dBm MAX
0 VDC
+27 dBm MAX
50 VDC
ALL RF PORTS
50
ESD
SENSITIVE
50 VDC MAX
REF IN
SOURCE 1
OUT
PORT 2
8
|
ni.com
|
NI PXIe-5632 Getting Started Guide