NI PXIe-4610 Calibration Procedure

|

© National Instruments Corporation

|

9

EEPROM Update

After completing a verification procedure, complete the following steps:
1.

Open a calibration session for the NI PXIe-4610. The default password is 

NI

.

2.

Store the calibration temperature.
a.

Record the ambient temperature.

b.

Call the DAQmx Set Temperature DSA Calibration function to store the calibration 
temperature measured in step a.

3.

Store the calibration constants.
a.

Perform steps 1 through 9 of the 

Gain and Residual Offset Verification

 section to 

calculate the 

ResidualOffset

 of each channel for which you want to update the offset 

calibration constant.

b.

Continue with steps 10 through 12 of the 

Gain and Residual Offset Verification

 

section to calculate the 

UnadjustedGain

 of each channel for which you want to update 

the gain calibration constant.

c.

Call the DAQmx Adjust DSA Power Amplifier Calibration function for the 
NI PXIe-4610 and configure it as shown in Table 6.

d.

Repeat steps a through c for each channel for which you want to store a calibration 
constant.

4.

Close the calibration session.

Accuracy Under Calibration Conditions

The following calibration specifications are valid for the following conditions:

Ambient temperature is 23 °C ± 5 °C.

PXI Express chassis fan speed is set to HIGH and has clean fan filters.

Empty slots have filler panels installed.

Note

The test limits listed in Table 4 and 5 are derived using the values in Table 

and 7, respectively.

Table 6.  

DAQmx Adjust DSA Power Amplifier Calibration Configuration

Physical Channel

Gain

Offset

Dev1/ch

x

The value obtained from step a.

The value obtained from step b.

Summary of Contents for PXIe-4610

Page 1: ...l Setup 5 Verification 5 Gain and Residual Offset Verification 6 EEPROM Update 9 Accuracy Under Calibration Conditions 9 World Wide Support and Services 10 Software NI DAQmx 9 8 added support for the...

Page 2: ...ifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Information about creating applications that use the NI DAQmx driver Lab...

Page 3: ...Gain and Residual Offset Verification If this instrument is unavailable use a digitizer capable of measuring 10 Vpk sine waves at 1 kHz with an amplitude uncertainty of 0 1 dB or less a linearity erro...

Page 4: ...onnecting the NI PXIe 4610 The NI PXIe 4610 provides connections for 2 channels Refer to Figure 1 for the connection pinout Figure 1 NI PXIe 4610 Pinout 1 OUT CH0 2 OUT CH0 3 OUT CH1 4 OUT CH1 5 IN CH...

Page 5: ...ime for all of the instruments and equipment according to the manufacturer instructions Ensure that the PXI PXI Express chassis fan speed is set to High that the fan filters are clean and that the emp...

Page 6: ...ctions 2 Configure the function generator as shown in Table 2 3 Enable the function generator 4 Enable the NI PXIe 4610 outputs Additional information about enabling the outputs is available in the NI...

Page 7: ...residual signal The residual offset is the average of the residual signal 10 Connect the DMM as shown in Figure 3 Figure 3 DMM Connections 11 Repeat steps 7 and 8 and record the amplitude of the acqu...

Page 8: ...ed are derived using the values in Table 7 20 Compare the recorded residual offset for each channel to the limits in Table 5 Note The test limits listed are derived using the values in Table 8 If the...

Page 9: ...rification section to calculate the UnadjustedGain of each channel for which you want to update the gain calibration constant c Call the DAQmx Adjust DSA Power Amplifier Calibration function for the N...

Page 10: ...most recent NI PXIe 4610 Specifications document at ni com manuals World Wide Support and Services The National Instruments website is your complete resource for technical support At ni com support y...

Reviews: