background image

© 2014 National Instruments Corporation. All rights reserved. 

374185A-01

May14

Refer to the 

NI Trademarks and Logo Guidelines 

at 

ni.com/trademarks

 for more information on National Instruments trademarks. Other 

product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National 
Instruments products/technology, refer to the appropriate location: 

Help»Patents

 in your software, the 

patents.txt

 file on your media, or the 

National Instruments Patents Notice

 at 

ni.com/patents

. You can find information about end-user license agreements (EULAs) and third-party 

legal notices in the readme file for your NI product. Refer to the 

Export Compliance Information 

at 

ni.com/legal/export-compliance

 

for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES 
NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR 
ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable 
limited rights and restricted data rights as set forth in FAR 52.227-14s, DFAR 252.227-7014, and DFAR 252.227-7015.

Note

For operational specifications, refer to the most recent 

NI PXIe-4610 

Specifications

 document at 

ni.com/manuals

.

World Wide Support and Services

The National Instruments website is your complete resource for technical support. At 

ni.com/

support

 you have access to everything from troubleshooting and application development 

self-help resources to email and phone assistance from NI Application Engineers.

Visit 

ni.com/services

 for NI Factory Installation Services, repairs, extended warranty, and 

other services.

Visit 

ni.com/register

 to register your National Instruments product. Product registration 

facilitates technical support and ensures that you receive important information updates from NI.

National Instruments corporate headquarters is located at 11500 North Mopac Expressway, 
Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For 
telephone support in the United States, create your service request at 

ni.com/support

 or dial 

1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the 
Worldwide Offices section of 

ni.com/niglobal

 to access the branch office websites, which 

provide up-to-date contact information, support phone numbers, email addresses, and current 
events.

Table 7.  

NI PXIe-4610 Gain Accuracy Under Calibration Conditions

Input Signal Source

Gain Accuracy (±dB)

Unadjusted

0.031

Adjusted*

0.011

*

Input signal source gain adjusted using the gain calibration constant stored in the EEPROM.

Table 8.  

NI PXIe-4610 Residual Offset Under Calibration Conditions

Residual Offset

±1 mV

Summary of Contents for PXIe-4610

Page 1: ...l Setup 5 Verification 5 Gain and Residual Offset Verification 6 EEPROM Update 9 Accuracy Under Calibration Conditions 9 World Wide Support and Services 10 Software NI DAQmx 9 8 added support for the...

Page 2: ...ifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Information about creating applications that use the NI DAQmx driver Lab...

Page 3: ...Gain and Residual Offset Verification If this instrument is unavailable use a digitizer capable of measuring 10 Vpk sine waves at 1 kHz with an amplitude uncertainty of 0 1 dB or less a linearity erro...

Page 4: ...onnecting the NI PXIe 4610 The NI PXIe 4610 provides connections for 2 channels Refer to Figure 1 for the connection pinout Figure 1 NI PXIe 4610 Pinout 1 OUT CH0 2 OUT CH0 3 OUT CH1 4 OUT CH1 5 IN CH...

Page 5: ...ime for all of the instruments and equipment according to the manufacturer instructions Ensure that the PXI PXI Express chassis fan speed is set to High that the fan filters are clean and that the emp...

Page 6: ...ctions 2 Configure the function generator as shown in Table 2 3 Enable the function generator 4 Enable the NI PXIe 4610 outputs Additional information about enabling the outputs is available in the NI...

Page 7: ...residual signal The residual offset is the average of the residual signal 10 Connect the DMM as shown in Figure 3 Figure 3 DMM Connections 11 Repeat steps 7 and 8 and record the amplitude of the acqu...

Page 8: ...ed are derived using the values in Table 7 20 Compare the recorded residual offset for each channel to the limits in Table 5 Note The test limits listed are derived using the values in Table 8 If the...

Page 9: ...rification section to calculate the UnadjustedGain of each channel for which you want to update the gain calibration constant c Call the DAQmx Adjust DSA Power Amplifier Calibration function for the N...

Page 10: ...most recent NI PXIe 4610 Specifications document at ni com manuals World Wide Support and Services The National Instruments website is your complete resource for technical support At ni com support y...