NI PXI-5663 Calibration Procedure
NI-RFSA supports programming the calibration procedures in LabVIEW, C, and
LabWindows
™
/CVI
™
. Refer to Table 1 for file locations.
For more information about calibration VIs and functions, install NI-RFSA and refer to the
NI RF Vector Signal Analyzers Help
, accessible through
Start»Programs»National
Instruments»NI RFSA»Documentation»NI RF Vector Signal Analyzers Help
.
This calibration procedure calibrates the NI PXIe-5663 RF vector signal analyzer as a single
device. To calibrate the NI 5601 or NI 5652, refer to their calibration procedures at
ni.com/
manuals
.
Documentation Requirements
You might find the following documentation helpful as you write the calibration procedure:
•
NI 5663 Specifications
•
NI 5661/5663/5663E RF Vector Signal Analyzers Getting Started Guide
•
NI RF Vector Signal Analyzers Help,
which includes LabVIEW and LabWindows/CVI
application programming references
These documents are installed with NI-RFSA. You also can find the latest versions of these
documents at
ni.com/manuals
.
Calibration Interval
NI recommends that you perform a complete calibration for the NI 5663 at least once a year.
Adjust the recommended calibration interval based on the measurement accuracy demands
of your application.
Table 1.
File Locations
File Name and Location
Description
<
IVI>\bin\niRFSA.dll
NI-RFSA instrument driver containing the entire NI-RFSA
API, including calibration functions.
<
LabVIEW>\instr.lib\
niRFSA\niRFSA.llb
LabVIEW VI library containing VIs for calling the
NI-RFSA calibration API. You can access calibration VIs
from the NI-RFSA calibration section of the LabVIEW
Functions palette.