NI PXI-5663 Calibration Procedure
|
© National Instruments
|
17
LO Output Power Verification
Table 10 shows the test limits.
1.
Disconnect the NI 5652 from the NI 5601 if it is connected.
2.
Connect the RF source to the power splitter input through the 3.5 mm-to-3.5 mm cable.
3.
Connect the power splitter reference output to power sensor A.
4.
Connect the other power splitter output to the NI 5663 LO IN connector through the
3.5 mm (m) to 3.5 mm (m) adapter.
5.
Connect a 50
Ω
terminator to the NI 5652 RF OUT connector.
6.
Connect power sensor B to the NI 5663 LO OUT connector through the SMA to SMA
right-angle adapter. Figure 5 shows the completed equipment setup.
Figure 5.
LO Output Power Equipment Setup
7.
Create a new session for the NI 5663 using the niRFSA Initialize VI or the
niRFSA_init
function.