CALIBRATION PROCEDURE
NI PXI-5663
This document contains instructions for writing a manual calibration procedure for the
NI PXI-5663 (NI 5663) RF vector signal analyzer. For more information about calibration, visit
ni.com/calibration
.
Contents
Software Requirements............................................................................................................. 1
Documentation Requirements .................................................................................................. 2
Calibration Interval................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 5
Calibration Procedures ............................................................................................................. 5
Verifying Reference Accuracy ................................................................................. 9
Absolute Accuracy Verification ............................................................................... 9
Average Noise Level Verification............................................................................ 14
LO Output Power Verification ................................................................................. 17
Non-Input-Related Spurs (Residual Spurs) Verification.......................................... 19
Sideband Spurs Verification..................................................................................... 21
Phase Noise Verification .......................................................................................... 25
Appendix A: Non-Input-Related Spurs Test Frequencies........................................................ 28
Where to Go for Support .......................................................................................................... 32
Software Requirements
To calibrate the NI 5663, you must install NI-RFSA version 2.3 or later and the NI Spectral
Measurements Toolkit version 2.4 or later on the calibration system.
Note
NI-RFSA version 2.1 does not include adjustment VIs.
You can download NI-RFSA from
ni.com/updates
.