NI PXI-5663 Calibration Procedure
|
© National Instruments
|
5
Test Conditions
Follow these guidelines to optimize the connections and the environment during calibration:
•
Maintain an ambient temperature of 23 °C ±5 °C.
•
Keep the relative humidity between 10% and 90%, noncondensing.
•
Ensure that the PXI/PXI Express chassis fan speed is set to HIGH, that the fan filters, if
present, are clean, and that the empty slots contain filler panels. For more information, refer
to the
Maintain Forced-Air Cooling Note to Users
document available at
ni.com/
manuals
.
•
(Optional) Clean the chassis fan filters at least every six months. Some chassis include fan
filters. Depending on the amount of chassis use and the ambient dust levels, filters might
require more frequent cleaning. If regular maintenance of dirty or clogged filters is not
possible, you can remove the filters to maintain adequate cooling.
•
Allow a warm-up time of at least 30 minutes after NI-RFSA is loaded and recognizes the
NI 5663. The warm-up time ensures that the measurement circuitry of the NI 5663 is at a
stable operating temperature.
•
Use an SMA torque wrench (5 in. · lb, Agilent 8710-1582 or equivalent) to tighten any
connection with an SMA connector.
•
Use a 3.5 mm torque wrench (8 in. · lb, Agilent 8710-1765 or equivalent) to tighten 3.5 mm
or 2.92 mm (K) connections without an SMA connector.
•
Lock all test equipment to the same reference frequency. Refer to the
NI 5663/5663E
Timing Configurations
topic in the
NI RF Vector Signal Analyzers Help
.
Calibration Procedures
The calibration process includes the following procedures:
1.
—Install the NI 5663 and configure it in Measurement & Automation Explorer
(MAX).
2.
—Characterize the response of the test system. Use the
measured response during the verification tests.
3.
—Measure the operation of the NI 5663. This procedure checks
whether the device is operating within its published specifications.
If the NI 5663 fails any of the verification tests, complete the following steps:
Note
Complete the verification for all NI 5663 specifications where as-found data
is required before performing any module calibrations.
1.
Module Calibration—Perform a calibration of the individual modules. Refer to the
following module calibration procedures at
ni.com/manuals
:
•
NI PXI-5650/5651/5652 Calibration Procedure
•
NI PXIe-5622 Calibration Procedure
•
NI PXIe-5601 Calibration Procedure