NI PXI-4110 Calibration Procedure
|
© National Instruments
|
9
To ensure the system has had adequate time to settle, wait one second after requesting a new
voltage before taking a measurement with the DMM and the NI 4110.
The voltage measurement test limits in Table 6 are derived from the voltage measurement
accuracy specifications plus an offset voltage.
Table 6.
NI 4110 Output Parameters and Test Limits for Voltage Measurement
Accuracy Verification
Channel
Iteration
Output (V)
Current Limit (A)
Test Limit (V)
0
1
0
0.5
±0.004000
2
1.5
±0.004750
3
3
±0.005500
4
4.5
±0.006250
5
6
±0.007000
1
1
0
±0.005000
2
5
±0.007500
3
10
±0.010000
4
15
±0.012500
5
20
±0.015000
2
1
0
±0.005000
2
-5
±0.007500
3
-10
±0.010000
4
-15
±0.012500
5
-20
±0.015000