Quick Start Guide
Phase Matrix PXI-1420 MW Downconverter Module
25
different levels of IF attenuation; the corresponding selected value of
IF attenuation is displayed in the entry box below the slide control. The
entry box value may be changed by highlighting the numbers and
entering the change or by using the up/down buttons to the left of the
entry box. Values can be entered from 0 dB to 31.5 dB in 0.5 dB steps.
IF ATTEN 2 (dB)
This slide control (along with an entry box) is
used to select the level of IF attenuation and may be changed by
pointing to the slide control and dragging it up or down to select
different levels of IF attenuation; the corresponding selected value of
IF attenuation is displayed in the entry box below the slide control. The
entry box value may be changed by highlighting the numbers and
entering the change or by using the up/down buttons to the left of the
entry box. Values can be entered from 0 dB to 31.5 dB in 0.5 dB steps.
Select INPUT Path
These selection buttons are used to select
between AUX IN and RF A IN.
•
When AUX IN (100 kHz to 425 MHz) is selected, an externally
generated IF signal can be input that bypasses the down conversion
process of the PXI-1420; the signal is directed to the IF A OUT
connector.
•
When RF A IN (2.75 GHz to 26.5 GHz) is selected, the signal, after
being down converted, is directed to the IF A OUT connector.