background image

©

 National Instruments Corporation

5

Getting Started with the NI PCI-8254R

8.

To provide power to your camera, connect an unused power connector from the ATX power supply 
on your computer to the Camera Power ATX connector on your NI 8254R, shown in Figure 1.

Figure 1 shows the information about placement of the connectors on the NI 8254R.

Figure 1.  

NI 8254R Connectors

9.

Replace the computer cover.

10. Connect the IEEE 1394a cable to your camera. Refer to your camera manufacturer documentation 

for specific instructions about how to connect the IEEE 1394a cable to your camera.

11. Connect the IEEE 1394a cable to an IEEE 1394a connector on the NI 8254R front panel. Refer to 

Figure 1 for the location of the IEEE 1394a connectors on the NI 8254R.

12. Plug in and power on the computer.

The NI 8254R is now installed and the camera is connected.

Wiring Isolated Output Power to the NI 8254R

This section describes how to connect the NI desktop power supply for the isolated outputs (part number 
778794-01) to the NI 8254R. For instructions on how to connect a separate main supply, refer to the 

Connecting to the Isolated Outputs Power Supply

 section.

Caution

Do 

not

 connect the NI 8254R isolated power to a source less than 5 VDC or greater than 

30 VDC. Doing so could damage the NI 8254R. 

1

IEEE 1394a Connectors

2

2-Position Isolated Outputs Power Connector

3

44-Pin D-SUB Connector

4

Camera Power ATX Connector

2

1

3

4

N114

Summary of Contents for NI VISION PCI-8254R

Page 1: ...PCI 8254R...

Page 2: ...ndows Vista NI Vision Acquisition Software 8 2 1 or later which includes the NI IMAQdx and NI IMAQ I O driver software Optional software for developing applications NI Vision Builder for Automated Ins...

Page 3: ...he LabVIEW Help Safety Information Caution The following paragraphs contain important safety information you must follow when installing and operating the device Do not operate the device in a manner...

Page 4: ...lse withstand voltage levels that commonly occur in electrical distribution systems The following is a description of Measurement Installation3 Categories Measurement Category I is for measurements pe...

Page 5: ...uter Store the NI 8254R in the antistatic package when not in use Installation The following instructions are for general installation Refer to the documentation provided by your computer manufacturer...

Page 6: ...o an IEEE 1394a connector on the NI 8254R front panel Refer to Figure 1 for the location of the IEEE 1394a connectors on the NI 8254R 12 Plug in and power on the computer The NI 8254R is now installed...

Page 7: ...Figure 3 b Reconnect the wire that was connected to the C position on the 4 position power connector to the C position on the 2 position isolated outputs power connector shown in Figure 3 Figure 2 Rem...

Page 8: ...power supply cable around the ferrite two times as shown in Figures 4 and 5 Figure 4 Installing a Ferrite Connector on the Power Supply Cable Figure 5 Closed Ferrite on the Power Supply Cable 4 Plug t...

Page 9: ...cturer specifications to terminate the wires leading to the 2 position power connector shown in Figure 7 Ferrules are available from Phoenix Contact part number 3200519 Figure 7 Wiring a Third Party P...

Page 10: ...orer MAX the National Instruments configuration utility to configure the IEEE 1394 camera connected to the NI 8254R for acquisition Refer to the Measurement Automation Explorer Help for NI IMAQdx for...

Page 11: ...Hz maximum load resistance 100 k Minimum pulse generated 100 s Reverse polarity protection Yes IEEE 1394 Number of ports 2 Speed 100 200 or 400 Mbps Power Requirements IEEE 1394 bus power1 18 W maximu...

Page 12: ...electrical equipment for measurement control and laboratory use EN 61326 EMC requirements Minimum Immunity EN 55011 Emissions Group 1 Class A CE C Tick ICES and FCC Part 15 Emissions Class A Note For...

Page 13: ...ts calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 787...