NI TB-9214 Calibration Procedure

c.

Start the task.

d.

Average the readings and record the value as CJC

temp0

.

e.

Clear the task.

8.

Set the thermocouple calibrator output to CJC

temp0

.

 

This ensures that during the 

thermocouple measurements, the input channel measures very close to 0 V, which 
eliminates gain errors in the NI 9214.

9.

Acquire and average samples.
a.

Create and configure an AI temperature thermocouple channel for the NI 9214 
according to Table 4.

b.

Configure the AI temperature thermocouple channel timing according to Table 5.

c.

Disable open thermocouple detection on the NI 9214.

d.

Start the task.

Note

Make sure the thermocouple calibrator and NI 9214 are completely settled at the 
desired temperature setting before acquiring data. 

Wait at least 300 ms for the Ectron 1140A to settle before acquiring data with the 
NI 9214.

Disregard the first six seconds of the acquisition to compensate for the open 
thermocouple settling time of the NI 9214.

e.

Average the readings for each channel. Record the average of the readings for each as 
Temp

ai

x

 

where 

x

 is the channel number.

f.

Clear the task.

10. Perform the following calculation using the recorded Temp

ai

x

 and CJC

temp0

 values.

Temp

aix

 - CJC

temp0

 = Temp

ErrorCh0

11. Compare the calculation result for each channel to the Upper Limit and Lower Limit values 

in Table 6.

12. Remove the NI TB-9214 from the NI 9214.

Table 4.  

NI 9214 Thermocouple Channel Configuration

Physical

Channels

Minimum

Value

Maximum

Vale

Units

Thermocouple

Type

Autozero

ai0:15

0

100

°C

E

Every Sample

Table 5.  

NI 9214 Thermocouple Timing Configuration

Timing

Mode

Sample

Mode

Samples Per

Channel

Rate (S/s)

High Resolution

Finite

36

1

Summary of Contents for NI TB-9214

Page 1: ...ts Software 1 Documentation 2 Test Equipment 3 Building the Barrier Strip Assembly 4 Test Conditions 5 Initial Setup 6 Verification 6 Terminal Block CJC Verification 6 Where to Go for Support 9 Softwa...

Page 2: ...ons and Specifications NI TB 9214 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW progra...

Page 3: ...Module Use an NI 9214 module that is verified to be within product specifications Chassis NI cDAQ 9178 BarrierStrip Assembly Thermocouple Wire TT E 24 SLE 100 E type SLE 24 AWG 20 to 30 meters Thermo...

Page 4: ...Crimp the chromel spade lugs to the chromel wire to create 16 positive spade lugs Crimp the constantan spade lugs to the constantan wire to create 16 negative spade lugs 2 Use the thermocouple wire sp...

Page 5: ...I TB 9214 meets published specifications Keep connections to the NI TB 9214 as short as possible Long cables and wires act as antennae picking up extra noise that can affect measurements Verify that a...

Page 6: ...er 6 Connect the power source to the NI cDAQ 9178 chassis 7 Launch Measurement Automation Explorer MAX 8 Right click the device name and select Self Test to ensure that the module is working properly...

Page 7: ...temp0 value a Create and configure an AI Temp Built In Sensor task for CJC channel _cjtemp0 on the NI 9214 according to Table 2 b Configure the AI Temp Built In Sensor task for CJC channel _cjtemp0 ac...

Page 8: ...fore acquiring data Wait at least 300 ms for the Ectron 1140A to settle before acquiring data with the NI 9214 Disregard the first six seconds of the acquisition to compensate for the open thermocoupl...

Page 9: ...at the NI TB 9214 is outside of the limits refer to Where to Go for Support for assistance in returning the device to NI Where to Go for Support The National Instruments Web site is your complete reso...

Reviews: